Enhancing benefits of bolus use through minimising the effect of air-gaps on dose distribution in photon beam radiotherapy

Published online by Cambridge University Press:  12 May 2020

Karim Bahhous*
Affiliation:
Department of Physics, Faculty of Science, Mohammed V University in Rabat, Rabat, Morocco Hassan II Oncology Center, University Hospital Mohammed VI, Oujda, Morocco
Mustapha Zerfaoui
Affiliation:
Department of Physics, Faculty of Science, University Mohamed 1st, Oujda, Morocco
Abdelaali Rahmouni
Affiliation:
Department of Physics, Faculty of Science Dhar El Mahraz, Sidi Mohamed Ben Abdellah University, Fez, Morocco
Naima El Khayati
Affiliation:
Department of Physics, Faculty of Science, Mohammed V University in Rabat, Rabat, Morocco
*
Author for correspondence: Karim Bahhous, Department of Physics, Faculty of Science, Mohammed V University in Rabat, Rabat, Morocco. E-mail: bahhouskarim@gmail.com
Rights & Permissions [Opens in a new window]

Abstract

Introduction:

Bolus material is frequently used on patient’s skin during radiation therapy to reduce or remove build-up effect for high-energy beams. However, the air-gaps formed between the bolus and the skin’s irregular surface reduce the accuracy of treatment planning. To achieve a good treatment outcome using bolus, experimental investigations are required to choose its thickness and to quantify the air-gap effect.

Material and methods:

Measurements for a 6 MV photon beam with a fixed source surface distance were carried out using the 31021 Semiflex 3D chamber into the water phantom. Firstly, the depth of maximum dose (R100) and the dose value at surface (Ds) were evaluated as a function of bolus thickness for some square fields. Secondly, to test the effect of the air-gaps ranged from 5 to 30 mm with a step of 5 mm between the bolus and the phantom surface, a water-equivalent RW3 (Goettingen White Water) slab form of 10 mm thickness was considered as a bolus.

Results:

We observed that the linear behaviour of R100 in terms of the bolus thickness makes the choice of this parameter more convenient depending on field size. In addition, increasing the air-gaps widens the penumbra and created electrons that have a greater probability to quit the radiation field borders before reaching the surface. The dose spread of the off-field area could have a significant influence on the patient treatment.

Conclusion:

Based on dose distribution comparisons between the measurements with and without air-gaps for the field size of 100 mm × 100 mm, it has been demonstrated that a maximum air-gap value lower than 5 mm would be desirable for an efficient use of the bolus technique.

Information

Type
Original Article
Copyright
© The Author(s), 2020. Published by Cambridge University Press
Figure 0

Figure 1. Experimental set-up used to quantify the air-gap effect on dose distribution using a slab and a water tank.

Figure 1

Figure 2. (a) PDD curves normalised at their own maximums for different bolus thicknesses measured into a water phantom. (b) Measured data points of R100 and Ds as a function of bolus thickness, the data points of R100 and Ds are fitted with linear function and logistic function, respectively. 6 MV photon beam and 100 mm × 100 mm field size.

Figure 2

Table 1. Fitting coefficients (and standard errors) obtained for linear (logistic) function applied to R100 (Ds) variation as a function of bolus thicknesses

Figure 3

Figure 3. Profiles shape variation in terms of air-gap thickness for 6 MV beams, 10 × 10 field size and a 10 mm bolus thickness. The curves are normalised to their own central dose value. Measured data at (a) phantom’s surface and at (b) 5 mm depth.

Figure 4

Figure 4. The gamma 2D index (a, c and e) and failed points (b, d and f) calculated by comparison between reference dose distribution and those with air-gaps. For a 6 MV beam, 100 mm × 100 mm field size with 30 mm of out-of-field and 10 mm bolus thickness. Acceptance criteria were defined to be ΔdM = 2 mm and ΔDM = 2%.

Figure 5

Table 2. Statistics of the gamma 2D index, the Local_Diff (%) and the DoN (%) obtained from comparison between reference dose distribution and those with air-gaps 5, 10, 15, 20, 25 and 30 mm, respectively

Figure 6

Table 3. Uncertainty budget for dose measurement