Post-market surveillance to detect adverse events associated with Melody® valve implantation

Published online by Cambridge University Press:  10 November 2016

Kevin D. Hill*
Affiliation:
Department of Pediatrics, Duke University Medical Center, Durham, North Carolina, United States of America
Bryan H. Goldstein
Affiliation:
The Heart Institute, Cincinnati Children’s Hospital Medical Center, Cincinnati, Ohio, United States of America
Michael J. Angtuaco
Affiliation:
Department of Pediatrics, University of Arkansas for Medical Sciences, Little Rock, Arkansas, United States of America
Patricia Y. Chu
Affiliation:
Department of Pediatrics, Duke University Medical Center, Durham, North Carolina, United States of America
Gregory A. Fleming
Affiliation:
Department of Pediatrics, Duke University Medical Center, Durham, North Carolina, United States of America
*
Correspondence to: K. D. Hill, MD, MS, Duke Clinical Research Institute, 2400 Pratt Street, Durham, NC 27705, United States of America. Tel: 919 668 8305; Fax: 1919 287 2612; E-mail: kevin.hill@duke.edu
Rights & Permissions [Opens in a new window]

Abstract

Objective

The aim of this study was to describe previously unrecognised or under-recognised adverse events associated with Melody® valve implantation.

Background

In rare diseases and conditions, it is typically not feasible to conduct large-scale safety trials before drug or device approval. Therefore, post-market surveillance mechanisms are necessary to detect rare but potentially serious adverse events.

Methods

We reviewed the United States Food and Drug Administration’s Manufacturer and User Facility Device Experience (MAUDE) database and conducted a structured literature review to evaluate adverse events associated with on- and off-label Melody® valve implantation. Adverse events were compared with those described in the prospective Investigational Device Exemption and Post-Market Approval Melody® transcatheter pulmonary valve trials.

Results

We identified 631 adverse events associated with “on-label” Melody® valve implants and 84 adverse events associated with “off-label” implants. The most frequent “on-label” adverse events were similar to those described in the prospective trials including stent fracture (n=210) and endocarditis (n=104). Previously unrecognised or under-recognised adverse events included stent fragment embolisation (n=5), device erosion (n=4), immediate post-implant severe valvar insufficiency (n=2), and late coronary compression (n=2 cases at 5 days and 3 months after implantation). Under-recognised adverse events associated with off-label implantation included early valve failure due to insufficiency when implanted in the tricuspid position (n=7) and embolisation with percutaneous implantation in the mitral position (n=5).

Conclusion

Post-market passive surveillance does not demonstrate a high frequency of previously unrecognised serious adverse events with “on-label” Melody® valve implantation. Further study is needed to evaluate safety of “off-label” uses.

Information

Type
Original Articles
Copyright
© Cambridge University Press 2016 
Figure 0

Figure 1 Melody® valve medical device reports identified from the United States Food and Drug Administration database and Melody® valve adverse event reports identified from the published literature.

Figure 1

Table 1 Melody® valve procedural adverse events.

Figure 2

Table 2 Melody® valve post-procedural adverse events.

Figure 3

Table 3 Case descriptions for previously under-recognised Melody® valve post-procedural adverse events.

Figure 4

Table 4 Adverse events associated with “off-label” Melody® valve implantation.