Volumetric and dosimetric comparison of computerised radiotherapy treatment plan between using positron emission tomography/computed tomography (PET/CT) and CT images for target delineation in non-small cell lung cancer patients

Published online by Cambridge University Press:  18 April 2016

Sanphat Sangudsup
Affiliation:
Medical Physics Master Degree Program, Department of Radiology, Faculty of Medicine, Chiang Mai University, Chiang Mai, Thailand
Tawika Kaewchur
Affiliation:
Department of Radiology, Division of Nuclear Medicine, Faculty of Medicine, Chiang Mai University, Chiang Mai, Thailand
Waralee Teeyasoontranon
Affiliation:
Department of Radiology, Division of Nuclear Medicine, Faculty of Medicine, Chiang Mai University, Chiang Mai, Thailand
Pitchayaponne Klunklin
Affiliation:
Division of Therapeutic Radiology and Oncology, Faculty of Medicine, Chiang Mai University, Chiang Mai, Thailand
Nisa Chawapun
Affiliation:
Division of Therapeutic Radiology and Oncology, Faculty of Medicine, Chiang Mai University, Chiang Mai, Thailand
Supoj Ua-apisitwong*
Affiliation:
Department of Radiology, Division of Nuclear Medicine, Faculty of Medicine, Chiang Mai University, Chiang Mai, Thailand
*
Correspondence to: Supoj Ua-apisitwong, Maharaj Nakorn Chiang Mai Hospital, Faculty of Medicine, Chiang Mai University, 239 Huay Kaew Road, Muang District, Chiang Mai, Thailand, 50200. Tel: +66 83 5668983. E-mail: apisitwong_sp@yahoo.com
Rights & Permissions [Opens in a new window]

Abstract

Purpose

To compare intensity-modulated radiation therapy (IMRT) treatment planning between using positron emission tomography/computed tomography (PET/CT) and CT for target volume delineation in patients with non-small cell lung cancer (NSCLC).

Methods

Nine NSCLC patients with PET/CT images were enrolled into this study. Gross tumour volumes (GTVs) were delineated by the PET visual assessment (PETvis), the automated PET (PETauto), standardised uptake value (SUV)>2·5 (PET2·5) and threshold 40% SUVmax (PET40), and CT-based method. For each patient, two IMRT treatment plans based on CT and PET/CT delineation were performed. The target coverage and the dose–volume parameters for organs at risk were analysed.

Results

The PETauto referred to PET40 when SUVmax<7 and PET2·5 when SUVmax≥7. The mean GTVs were 15·04, 15·7 and 15·14 cc for PETauto, PETvis and CT based, respectively. The GTV of PETauto was not different from PETvis (p=0·441) and CT based (p=0·594). Based on CT delineation in IMRT planning, only 34% of the cases had sufficient PET/CT planning target volumes coverage, whereas the organs at risk dose parameters were not statistically significant (p>0·05).

Conclusions

PET/CT enables more accurate assessment of tumour delineation for NSCLC, therefore improve target coverage in IMRT plan.

Information

Type
Original Articles
Copyright
© Cambridge University Press 2016 
Figure 0

Figure 1 Relationship of the percentage difference between the gross tumour volumes (GTVs) contoured, based on positron emission tomography visual assessment (PETvis) versus PET40 (red triangle), PETvis versus PET2·5 (black dot) compared with maximum standardised uptake value (SUV) of tumour. The vertical axis indicates the percentage GTV difference between the two methods.

Figure 1

Table 1 Results of gross tumour volume (GTV), distance of GTV centre point and conformity index between positron emission tomography automated (PETauto) and PET visual assessment (PETvis)

Figure 2

Table 2 Results of gross tumour volume (GTV), distance of GTV centre point and conformity index between positron emission tomography automated (PETauto) and computed tomography (CT) based

Figure 3

Figure 2 Coverage of positron emission tomography/computed tomography (PET/CT) planning target volume (PTV) derived from CT planning.

Figure 4

Figure 3 Example computed tomography (CT) slide of insufficient target coverage of positron emission tomography/CT (PET/CT) planning target volume (PTV) derived from CT planning in patient no. 7.

Figure 5

Table 3 Dose–volume histogram parameters for organs at risk of plan computed tomography (CT) and plan positron emission tomography/CT (PET/CT)