Coronary compression testing by balloon interrogation during pulmonary valve implantation: room for doubt?

Published online by Cambridge University Press:  10 February 2021

Gareth J. Morgan*
Affiliation:
The Heart Institute, Children’s Hospital Colorado, Aurora, CO, USA University of Colorado, School of Medicine, Anschutz Medical Campus, Aurora, CO, USA Department of Cardiology, University of Colorado Hospital, Aurora, CO, USA
Salvador A. Rodriguez
Affiliation:
The Heart Institute, Children’s Hospital Colorado, Aurora, CO, USA University of Colorado, School of Medicine, Anschutz Medical Campus, Aurora, CO, USA
Ryan Leahy
Affiliation:
The Heart Institute, Children’s Hospital Colorado, Aurora, CO, USA University of Colorado, School of Medicine, Anschutz Medical Campus, Aurora, CO, USA
Jenny E. Zablah
Affiliation:
The Heart Institute, Children’s Hospital Colorado, Aurora, CO, USA University of Colorado, School of Medicine, Anschutz Medical Campus, Aurora, CO, USA
*
Author for correspondence: Gareth J Morgan, MD, The Heart Institute, Children’s Hospital Colorado, 13123 East 16th Avenue, Box 100, Aurora, CO 80045, USA. Tel: +17207773379; Fax: +17207773374. E-mail: drgarethjmorgan@gmail.com
Rights & Permissions [Opens in a new window]

Abstract

Objective:

To evaluate the reliability of balloon coronary compression testing during percutaneous pulmonary valve implantation.

Background:

Despite the widespread use of the ‘balloon coronary test’ as the preferable method to rule out the risk of coronary compression, this adverse event has been described after pulmonary valve implantation where coronary balloon test suggested no risk or low risk, calling into question the accuracy of the test.

Methods:

We performed a retrospective chart review of 84 patients who underwent pulmonary valve implantation between January 2018 and December 2019 and selected 36 patients whose archived imaging was suitable to perform quantitative analysis of the ‘balloon coronary test’. We focused on the spatial disparity between the right ventricular outflow tract position defined by the inflated testing balloon and the eventual implanted valve position, to classify the test as inaccurate or accurate.

Results:

In total, 36.1% of cases were classified as having an inaccurate coronary balloon test. Among the baseline characteristics, right ventricular outflow tract substrate was identified as a significant predictor of test accuracy. Related to this characteristic, the type of testing balloon used and the size of the eventually implanted valve were found to be associated with test accuracy.

Conclusions:

Based on our findings, balloon coronary testing is not an accurate method of predicting final valve position with respect to fixed structures in the thorax. This may translate to a high false positive rate for the likelihood of coronary compression in pulmonary valve implantation.

Information

Type
Original Article
Copyright
© The Author(s), 2021. Published by Cambridge University Press
Figure 0

Figure 1. Distance between the posterior border of the right ventricular outflow tract and the sternum. During the quantitative assessment three measurements (1, 2, 3) were performed to establish the difference between the right ventricular outflow tract posterior border at T1 (a) and T2 (b). The location of these measurements was defined based on the overlaid image (c).

Figure 1

Figure 2. Posterior position of the testing balloon compared to the valve. In these examples, we have overlaid the coronary compression balloon test with the final valve image at the same angiographic angle. In each case, you can easily see that the balloon position is more posteriorly positioned that the final valve position. It is not clear how each individual part of the heart (specifically, the right ventricular outflow tract and the aortic root and coronary arteries) moves during balloon testing; therefore, we limited our analysis to determine the relative position of the posterior border of the right ventricular outflow tract [posterior border of the balloon (white-dashed line) or valve frame] with respect to the posterior wall of the sternum.

Figure 2

Table 1. Baseline and procedural characteristics

Figure 3

Table 2. Measurements between the posterior border of right ventricular outflow tract and the sternum.

Figure 4

Figure 3. Distribution of the observed difference between the right ventricular outflow tract posterior wall at T1 and T2. (a) Comparison between the distribution observed in our population (dark line) and the ideal distribution expected for this test (grey line). (b). Posterior position of the balloon compared with the final position of the valve.