Comparison of verification accuracy and radiation dose between megavoltage CT and kilovoltage cone-beam CT

Published online by Cambridge University Press:  12 November 2010

Vincent W.C. Wu
Affiliation:
Department of Health Technology and Informatics, Hong Kong Polytechnic University, Hung Hom, Kowloon, Hong Kong
M.L. Ho
Affiliation:
Department of Health Technology and Informatics, Hong Kong Polytechnic University, Hung Hom, Kowloon, Hong Kong
L.Y. Yuen
Affiliation:
Department of Health Technology and Informatics, Hong Kong Polytechnic University, Hung Hom, Kowloon, Hong Kong
K.L. Li
Affiliation:
Department of Health Technology and Informatics, Hong Kong Polytechnic University, Hung Hom, Kowloon, Hong Kong
N.F. Lai
Affiliation:
Department of Health Technology and Informatics, Hong Kong Polytechnic University, Hung Hom, Kowloon, Hong Kong
Y.C. Tsang
Affiliation:
Department of Health Technology and Informatics, Hong Kong Polytechnic University, Hung Hom, Kowloon, Hong Kong
W.Y. Lee
Affiliation:
Department of Radiotherapy and Oncology, Hong Kong Sanatorium and Hospital, Happy Valley, Hong Kong
G. Chiu
Affiliation:
Department of Radiotherapy and Oncology, Hong Kong Sanatorium and Hospital, Happy Valley, Hong Kong
Rights & Permissions [Opens in a new window]

Abstract

Purpose: This study evaluated the difference in verification performance and organ doses between megavoltage computed tomography (MVCT) and kilovoltage cone-beam computed tomography (kV CBCT).

Methods: Anthropomorphic phantoms of head-and-neck (H&N) and pelvic regions were scanned by the computed tomography–simulator. A common reference standard setup position for each phantom was determined. Markings of known deviations from the standard position in the lateral, longitudinal and yaw displacements were made on the phantoms. The verifications by kV CBCT and MVCT were conducted in the linear accelerator and helical tomotherapy treatment unit respectively. The phantoms were then shifted in steps according to the assigned different degree of positional deviations. The discrepancy between the detected data and the known data (δD) indicated the detection error of the verification system. Thermoluminescent dosimeters were used for dose measurements of kV CBCT and MVCT.

Results: δD of the lateral translations and yaw rotation in kV CBCT of the H&N region were lower than MVCT. There were no differences in the error detection along longitudinal direction between the two verification methods. For H&N, the mean doses of the various organs were significantly lower in kV CBCT except the skin. For the pelvis, MVCT delivered significantly higher mean dose to the prostate and femoral heads.

Information

Type
Original Article
Copyright
Copyright © Cambridge University Press 2011
Figure 0

Figure 1. The anthropomorphic head-and-neck phantom showing the cross-section at the mid-neck level. The circular spots are the possible locations for the placement of thermoluminescent dosimeter (some of the holes have been un-plagued for illustration purpose).

Figure 1

Figure 2. Schematic diagram showing the assigned deviation lines in the x and y directions marked on the phantom (not to scale).

Figure 2

Figure 3. Schematic diagram showing the assigned deviation lines in yaw rotation marked on the phantom (not to scale).

Figure 3

Table 1. Comparison of δD (discrepancy between the measured data and the known data) in the x direction between kV CBCT and MVCT

Figure 4

Table 2. Comparison of δD (discrepancy between the measured data and the known data) in the y direction between kV CBCT and MVCT

Figure 5

Table 3. Comparison of δD (discrepancy between the measured data and the known data) in the yaw rotation between kV CBCT and MVCT

Figure 6

Table 4. Comparison of organ doses per scan in the head-and-neck region between kV CBCT and MVCT

Figure 7

Table 5. Comparison of organ doses per scan in the pelvis region between kV CBCT and MVCT

Figure 8

Table 6. Comparison of estimated total organ doses received during a full course of radiotherapy (35 fractions) between kV CBCT and MVCT