Compatibility of Atmospheric 14CO2 Measurements: Comparing the Heidelberg Low-Level Counting Facility to International Accelerator Mass Spectrometry (AMS) Laboratories

Published online by Cambridge University Press:  19 September 2016

Samuel Hammer*
Affiliation:
Institut für Umweltphysik, Heidelberg University, Germany
Ronny Friedrich
Affiliation:
Curt Engelhorn Center for Archaeometry gGmbH, Mannheim, Germany
Bernd Kromer
Affiliation:
Institut für Umweltphysik, Heidelberg University, Germany Curt Engelhorn Center for Archaeometry gGmbH, Mannheim, Germany
Alexander Cherkinsky
Affiliation:
Center for Applied Isotope Studies, University of Georgia, USA
Scott J Lehman
Affiliation:
INSTAAR, University of Colorado, Boulder, Colorado, USA
Harro A J Meijer
Affiliation:
Centre for Isotope Research (CIO), Energy and Sustainability Research Institute Groningen (ESRIG), University of Groningen, the Netherlands
Toshio Nakamura
Affiliation:
Center for Chronological Research, Nagoya University, Japan
Vesa Palonen
Affiliation:
Department of Physics, University of Helsinki, Finland
Ron W Reimer
Affiliation:
14CHRONO Centre for Climate, the Environment and Chronology,School of Geography, Archaeology and Palaeoecology, Queen’s University Belfast, UK
Andrew M Smith
Affiliation:
Australian Nuclear Science and Technology Organisation, Lucas Heights, NSW 2234, Australia
John R Southon
Affiliation:
Earth System Science Department, University of California, Irvine, California 92612, USA
Sönke Szidat
Affiliation:
Department of Chemistry and Biochemistry & Oeschger Centre for Climate Change Research, University of Bern, Switzerland
Jocelyn Turnbull
Affiliation:
National Isotope Centre, GNS Science New Zealand and CIRES, University of Colorado, USA
Masao Uchida
Affiliation:
National Institute for Environmental Studies, Tsukuba, Japan
*
*Corresponding author. Email: shammer@iup.uni-heidelberg.de.
Rights & Permissions [Opens in a new window]

Abstract

Combining atmospheric Δ14CO2 data sets from different networks or laboratories requires secure knowledge on their compatibility. In the present study, we compare Δ14CO2 results from the Heidelberg low-level counting (LLC) laboratory to 12 international accelerator mass spectrometry (AMS) laboratories using distributed aliquots of five pure CO2 samples. The averaged result of the LLC laboratory has a measurement bias of –0.3±0.5‰ with respect to the consensus value of the AMS laboratories for the investigated atmospheric Δ14C range of 9.6 to 40.4‰. Thus, the LLC measurements on average are not significantly different from the AMS laboratories, and the most likely measurement bias is smaller than the World Meteorological Organization (WMO) interlaboratory compatibility goal for Δ14CO2 of 0.5‰. The number of intercomparison samples was, however, too small to determine whether the measurement biases of the individual AMS laboratories fulfilled the WMO goal.

Information

Type
Advances in Physical Measurement Techniques
Copyright
© 2016 by the Arizona Board of Regents on behalf of the University of Arizona 
Figure 0

Table 1 Participating laboratories in alphabetical order, which is not identical to the lab number used in this study.

Figure 1

Table 2 Summary of the ICP samples.

Figure 2

Table 3 Remarks and problems with the ICP samples as reported by the laboratories.

Figure 3

Table 4 Measurement bias (i.e. mean difference of measured Δ14C minus consensus value for all five CO2 samples) and respective uncertainties given as standard error of the mean. In addition, the reduced χ² values are listed, based on the consensus value and the median (see text for explanation). Results from laboratories marked by an asterisk have not been included in the calculation of the consensus value.

Figure 4

Figure 1 Differences of the individual labs to the consensus value of sample 30874 (OxI). The gray shaded area indicates the uncertainty of the consensus value. The difference between the consensus value and the nominal value of the NIST oxalic acid I (SRM 4990B) is shown as red dashed line. Lab 6 reported insufficient graphitization for this sample 30874.

Figure 5

Figure 2 Summary of all ICP results. The difference for each sample to the consensus value based on 9 labs is shown. Labs 3, 6, and 11 have been excluded from calculation of the consensus value (compare also reduced χ2 vs. the median in Table 4 for those labs). The measurements in brackets from labs 5 and 6 are subject to sample handling problems (compare Table 3).