Evaluating the dosimetric consequences of MLC leaf positioning errors in dynamic IMRT treatments

Published online by Cambridge University Press:  12 February 2019

Arpita Agarwal*
Affiliation:
Department of Physics, School of Sciences, IFTM University, Moradabad, Uttar Pradesh, India
Nikhil Rastogi
Affiliation:
Department of Physics, School of Sciences, IFTM University, Moradabad, Uttar Pradesh, India
KJ Maria Das
Affiliation:
Department of Radiotherapy, Sanjay Gandhi Postgraduate Institute of Medical Sciences, Lucknow, Uttar Pradesh, India
SA Yoganathan
Affiliation:
Department of Radiotherapy, Sanjay Gandhi Postgraduate Institute of Medical Sciences, Lucknow, Uttar Pradesh, India
D Udayakumar
Affiliation:
Department of Radiotherapy, Sanjay Gandhi Postgraduate Institute of Medical Sciences, Lucknow, Uttar Pradesh, India
R Naresh
Affiliation:
Department of Radiotherapy, Sanjay Gandhi Postgraduate Institute of Medical Sciences, Lucknow, Uttar Pradesh, India
Shaleen Kumar
Affiliation:
Department of Radiotherapy, Sanjay Gandhi Postgraduate Institute of Medical Sciences, Lucknow, Uttar Pradesh, India
*
Author for correspondence: Ms. Arpita Agarwal, Department of Physics, School of Sciences, IFTM University, Moradabad, Uttar Pradesh, India. Tel: 12253605730. E-mail: arpitagrwl@gmail.com
Rights & Permissions [Opens in a new window]

Abstract

Purpose

The purpose of this study was to evaluate the dosimetric impact of multileaf collimator (MLC) positional errors on dynamic intensity-modulated radiotherapy (IMRT) treatments through planning simulation. Secondly the sensitivity of IMRT MatriXX device for detecting the MLC leaf positional errors was also evaluated.

Materials and methods

In this study five dynamic IMRT plans, each for brain and head–neck (HN), were retrospectively included. An in-house software was used to introduce random errors (uniform distribution between −2·0 and +2·0 mm) and systematic errors [±0·5, ±0·75, ±1·0 and ±2·0 mm (+: open MLC error and −: close MLC error)]. The error-introduced MLC files were imported into the treatment planning system and new dose distributions were calculated. Furthermore, the dose–volume histogram files of all plans were exported to in-house software for equivalent uniform dose (EUD), tumour control probability and normal tissue complication probability calculations. The error-introduced plans were also delivered on LINAC, and the planar fluences were measured by IMRT MatriXX. Further, 3%/3 mm and 2%/2 mm γ-criteria were used for analysis.

Results

In planning simulation study, the impact of random errors was negligible and ΔEUD was <0·5±0·7%, for both brain and HN. The impact of systematic errors was substantial, and on average, the maximum change in EUD for systematic errors (close 2 mm) was −10·7±3·1% for brain and −15·5±2·6% for HN.

Conclusions

It can be concluded that the acceptable systematic error was 0·4 mm for brain and 0·3 mm for HN. Furthermore, IMRT MatriXX device was able to detect the MLC errors ≥2 mm in HN and >3 mm errors in brain with 2%/2 mm γ-criteria.

Information

Type
Original Article
Copyright
© Cambridge University Press 2019 
Figure 0

Table 1 The values of different parameters used in brain and HN for calculating biological parameters

Figure 1

Figure 1 DVH of CTV for all plans of one brain case. Abbreviations: DVH, dose–volume histogram; CTV, clinical target volume.

Figure 2

Table 2 Average values of physical parameters (D95 for CTV and Dmax for OARs) (mean ± SD) (Gy) for all five brain patients

Figure 3

Table 3 Average EUD (mean ± SD) (Gy) for all five brain patients

Figure 4

Table 4 Percentage of pixels passing gamma test (mean±SD) for brain patients

Figure 5

Figure 2 (a,b) DVH of CTV1 and CTV2 for all plans of one HN case. Abbreviations: DVH, dose–volume histogram; CTV, clinical target volume; HN, head–neck.

Figure 6

Table 5 Average values of physical parameters (mean ± SD) (Gy) for all five HN patients

Figure 7

Table 6 Average EUD (mean ± SD) (Gy) for all five HN patients

Figure 8

Table 7 Percentage of pixels passing gamma test (mean±SD) for HN patients

Figure 9

Figure 3 Changes in target EUD due to systematic MLC leaf errors in brain (CTV) and HN (CTV1 and CTV2).

Figure 10

Table 8 Acceptable systematic MLC leaf errors (mm) for brain

Figure 11

Table 9 Acceptable systematic MLC leaf errors (mm) for HN

Figure 12

Table 10 Impact of actual MLC leaf errors on brain cases

Figure 13

Table 11 Impact of actual MLC leaf errors on HN cases