Hostname: page-component-74656f9884-c52kr Total loading time: 0 Render date: 2026-11-07T11:20:35.544Z Has data issue: false hasContentIssue false

Preclinical studies for implementing flattening filter-free beams in hypofractionated volumetric-modulated arc therapy (VMAT) for prostate cancer radiotherapy

Published online by Cambridge University Press:  01 September 2017

Baochang Liu
Affiliation:
Department of Medical Physics, British Columbia Cancer Agency, Centre for the North, Prince George, BC, Canada
Johnson Darko
Affiliation:
Department of Medical Physics, Grand River Regional Cancer Center, Kitchener, ON, Canada Department of Physics and Astronomy, University of Waterloo, Waterloo, ON, Canada Department of Clinical Studies, Ontario Veterinary College, University of Guelph, Guelph, ON, Canada
Ernest Osei*
Affiliation:
Department of Medical Physics, Grand River Regional Cancer Center, Kitchener, ON, Canada Department of Physics and Astronomy, University of Waterloo, Waterloo, ON, Canada Department of Clinical Studies, Ontario Veterinary College, University of Guelph, Guelph, ON, Canada Department of Systems Design, University of Waterloo, Waterloo, ON, Canada
*
Correspondence to: Ernest Osei, Department of Medical Physics, Grand River Regional Cancer Centre, 835 King Street West, Kitchener, ON, Canada N2G1G3. Tel: +1 519 749 4300, ext 5407. E-mail: ernest.osei@grhosp.on.ca
Rights & Permissions [Opens in a new window]

Abstract

Background

Volumetric-modulated arc therapy (VMAT) has emerged as one of the most favourable techniques for radiotherapy treatment in recent years because of its conformal dose distribution to the planning target volume (PTV), lower doses to adjacent normal organs at risk (OARs) and faster and easier dose delivery. A typical conventional VMAT protocol for low-intermediate risk prostate cancer uses a flattened 6 MV photon beam to deliver 78 Gy in 39 fractions, however, a recent Radiation Therapy Oncology Group study investigated prostate cancer radiotherapy with a hypofractionated dose scheme of 36·25 Gy in 5 fractions. One advantage of flattening filter-free (FFF) beams in radiotherapy is the higher doses in the central region on the dose profile and much higher dose delivery rates.

Methods and materials

This paper reports the investigation of preclinical studies for implementing FFF beams in hypofractionated VMAT for prostate cancer radiotherapy. All treatment planning were accomplished using Varian EclipseTM treatment planning system version 11 and delivered on Varian Truebeam linear accelerators. The studies compared the biological-effective dose–volume histograms and dose–volume histograms of PTV and OARs for 20 patients using conventional and hypofractionated dose schemes. The study also evaluated the 6 and 10 MV FFF by comparing 6 and 10 MV VMAT plans with the FFF beams. The treatment time was investigated using plans with 6 MV beams and doses of 2, 4, 5, 6, 7·25 Gy/fraction and plans with 10 MV FFF with a dose of 7·25 Gy/fraction. We also investigated an angular monitor unit (MU) quantity (MU/deg) and its threshold value for RapidArcTM plans, beyond which FFF beams can be considered superior to flattened beams in terms of treatment time increased caused by higher dose per fraction.

Results

The results show that the hypofractionated plans resulted in greater biological equivalent doses to PTV and lower doses to OARs. The 10 MV FFF plans have statistically lower mean doses to all the OARs, whereas PTV homogeneity index remains the same compared with other beam energies. The mean body integral dose for the 20 patients is 8·7% lower using 10 MV FFF compared with 6 MV FFF mainly because of the higher energy and less required MUs with the 10 MV FFF beam. The hypofractionated scheme with 10 MV FFF plan has the same treatment time as that of the 6 MV plan at 2 Gy/fraction, as the higher dose delivery rates at 10 MV FFF can compensate for the higher prescribed dose per fraction without the need of extra treatment time.

Conclusion

In this study, we observed that the 10 MV FFF beam is better for hypofractionated prostate cancer VMAT plan delivery. The threshold value of MU/deg is found to be 2·083 MU/deg based on our machine configurations.

Information

Type
Original Articles
Copyright
© Cambridge University Press 2017 
Figure 0

Table 1 Average and 1 SD of both the mean doses to organs at risk and planning target volume (PTV) homogeneity index (HI) for all 20 patients

Figure 1

Figure 1 Comparison of average organs at risk dose–volume histograms (DVHs): 6 MV versus 6 MV flattening filter free (FFF): bladder (a), rectum (b), left femur (c) and right femur (d). Notes: The solid lines are the mean DVH and the ‘cloud of DHVs’ represents 1 SD above and below the mean.

Figure 2

Figure 2 Comparison of average organs at risk dose–volume histograms (DVHs): 6 MV versus 10 MV flattening filter free (FFF): bladder (a), rectum (b), left femur (c) and right femur (d). Notes: The solid lines are the mean DVH and the ‘cloud of DHVs’ represents 1 SD above and below the mean.

Figure 3

Figure 3 Comparison of average organs at risk dose–volume histograms (DVHs): 6 versus10 MV: bladder (a), rectum (b), left femur (c) and right femur (d). Notes: The solid lines are the mean DVH and the ‘cloud of DHVs’ represents 1 SD above and below the mean.

Figure 4

Figure 4 Comparison of average organs at risk dose–volume histograms (DVHs): 10 versus 6 MV flattening filter free (FFF): bladder (a), rectum (b), left femur (c) and right femur (d). Notes: The solid lines are the mean DVH and the ‘cloud of DHVs’ represents 1 SD above and below the mean.

Figure 5

Figure 5 Comparison of average organs at risk dose–volume histograms (DVHs): 10 MV flattening filter free (FFF) versus 10 MV: bladder (a), rectum (b), left femur (c) and right femur (d). Notes: The solid lines are the mean DVH and the ‘cloud of DHVs’ represents 1 SD above and below the mean.

Figure 6

Figure 6 Mean planning target volume (PTV) dose–volume histograms (DVHs) of the 20 patients for 6 MV, 6 MV flattening filter free (FFF), 10 MV and 10 MV FFF beam energies. Notes: The solid lines are the mean DVH and the ‘cloud of DHVs’ represents 1 SD above and below the mean. However, the results closely overlap and are not apparent. Abbreviation: PTV_PROS, PTV prostate.

Figure 7

Figure 7 Comparison of average planning target volume (PTV) and organs at risk dose–volume histograms (DVHs) of hypofractionated volumetric-modulated arc therapy (VMAT) plans between 6 MV and 10 MV flattening filter free (FFF). (a) PTV_prostate, (b) bladder, (c) rectum, (d) left femur and (e) right femur. Notes: The solid lines are the mean DVH and the ‘cloud of DHVs’ represents 1 SD above and below the mean.

Figure 8

Table 2 The average and 1 SD of the mean doses to the organs at risk and of planning target volume (PTV) homogeneity index (HI) for the 20 patients

Figure 9

Figure 8 Comparison of average planning target volume (PTV) and organs at risk dose–volume histograms (DVHs) between conventional (6X, 10X FFF) and hypofractionated (Hypo 10X FFF, Hypo 6X) dose schemes at energies of 6 MV and 10 MV flattening filter free (FFF). (a) PTV_prostate, (b) bladder, (c) rectum, (d) left femur and (e) right femur.

Figure 10

Figure 9 Biological effective dose–volume histograms comparison between conventional (6X, 10X FFF) and hypofractionated (Hypo 10X FFF, Hypo 6X) dose schemes for 6 MV and 10 MV flattening filter free (FFF). (a) PTV_prostate, (b) bladder, (c) rectum, (d) left femur and (e) right femur. Abbreviation: BED, biological equivalent doses.

Figure 11

Figure 10 (a) Relationship between gantry rotational speed and monitor unit (MU) delivered per gantry angle for full counterclockwise (CCW) arcs with a range of MUs (288, 603, 782, 882, 921, 1,082 MU). (b) Relation between normalised gantry rotation time (t second/75 second) and total MU per arc for a range of prescribed fractional doses [2, 4, 5, 6 and 7·25 Gy/fraction at 6 MV and 7·25 Gy/fraction at 10 MV flattening filter free (FFF)]. Notes: The time of 75 seconds is the total time for one arc when gantry rotates at its maximum speed of 4·8 deg/second.

Figure 12

Figure 11 Deviation of the doses between measured by an ion chamber at the central cavity of the ArcCHECK device and calculated in the Eclipse treatment planning system. Abbreviation: FFF, flattening filter free.

Figure 13

Figure 12 Gamma evaluation of the plan quality assurance measurement using ArcCHECK. Notes: The prefix ‘CW’ means the clockwise arc and the prefix ‘CCW’ refers to the counterclockwise arc. The CW and CCW arcs were measured separately. The prefix ‘Total’ refers to both the CW and CCW arcs measured together. (a) Gamma evaluation criteria: 3 mm and 3%. (b) Gamma evaluation criteria: 2 mm and 2%. Abbreviation: FFF, flattening filter free.

Figure 14

Figure 13 Gamma evaluation of conventional plan quality assurance measurement using electronic portal imaging device. Note: The prefix ‘CW’ means the clockwise arc and the prefix ‘CCW’ refers to the counterclockwise arc. (a) Gamma evaluation criteria of the 3 mm and 3%. (b) Gamma evaluation criteria of the 2 mm and 2%. Abbreviation: FFF, flattening filter free.

Figure 15

Figure 14 Gamma evaluation of hypofractionated plan quality assurance measurements using electronic portal imaging device. Note: The prefix ‘CW’ means the clockwise arc and the prefix ‘CCW’ refer to the counterclockwise arc. (a) Gamma evaluation criteria: 3 mm and 3%. (b) Gamma evaluation criteria: 2 mm and 2%. Abbreviation: FFF, flattening filter free.