Opportunities and challenges surrounding the use of implantable haemodynamic monitor in paediatric Fontan patients living in high-altitude, rural settings

Published online by Cambridge University Press:  29 August 2025

Deepti Pagare Bhat*
Affiliation:
Division of Cardiology and multi-disciplinary Fontan clinic, Phoenix Children’s’ Hospital, Phoenix, AZ, USA University of Arizona, Phoenix, AZ, USA
Arash A. Sabati
Affiliation:
Division of Cardiology and multi-disciplinary Fontan clinic, Phoenix Children’s’ Hospital, Phoenix, AZ, USA University of Arizona, Phoenix, AZ, USA
Saleem Almasarweh
Affiliation:
Division of Cardiology and multi-disciplinary Fontan clinic, Phoenix Children’s’ Hospital, Phoenix, AZ, USA University of Arizona, Phoenix, AZ, USA
Joseph N. Graziano
Affiliation:
Division of Cardiology and multi-disciplinary Fontan clinic, Phoenix Children’s’ Hospital, Phoenix, AZ, USA University of Arizona, Phoenix, AZ, USA
*
Corresponding author: Deepti Pagare Bhat; Email: dbhat@arizona.edu
Rights & Permissions [Opens in a new window]

Abstract

Background:

Implantable haemodynamic monitors allow remote monitoring of Fontan circulation. We report unique opportunities and challenges related to device use in rural, high-altitude regions.

Objectives:

Assess the performance of implantable haemodynamic monitor in Fontan circulation and identify potential sources of measurement discrepancy defined as non-physiological, negative, or significantly lower reading than baseline.

Methods:

We performed a retrospective review of patients who underwent implantable haemodynamic monitor implantation from September 2021 to April 2024 (n = 17) at our centre (∼1,000 feet above sea level; ASL) and identified those with sensor discrepancies.

Results:

During a mean follow-up duration of 26 months (range 13–44 months), there were no procedure-related complications, thromboembolism, or device displacement. Ten patients lived in rural, higher-altitude regions (average altitude 5100 feet above sea level, average distance from centre ∼160 miles, range = 100–400 miles). Challenges in remote monitoring included unreliable home-internet connection, non-compliance, and difficulty performing device recalibration at patient’s home altitude. Sensor discrepancies were noted in 7 patients (41%), of whom 6 (86%) lived remotely. Manual review of the waveforms identified sources of discrepancy, including misinterpretation of the non-pulsatile pressure waveform (n = 3), offset due to change in hospital-interrogation unit (n = 4), and sensor drift (n = 1). Altitude change did not directly affect sensor performance. We were able to apply corrective interventions in 4/7 sensors, including Fontan-specific settings (overriding pulsatility), and back-end recalibration, which were effective in improving device accuracy.

Conclusions:

Implantable haemodynamic monitors are a promising tool for monitoring Fontan circulation but may require modified settings and careful attention to potential interpretation errors. Home monitoring remains challenging for rural, high-altitude residents with limited resources.

Information

Type
Original Article
Copyright
© The Author(s), 2025. Published by Cambridge University Press
Figure 0

Figure 1. (a) Technical considerations while implanting IHM in a Fontan patient. The left panel shows the target implant site in the branch left pulmonary artery. The right panel shows the fluoroscopic view of the implanted IHM in a Fontan patient with a pre-existing stent in the left pulmonary artery. (b) shows examples of sensor readings from Fontan patients.

Figure 1

Table 1. Demographic and haemodynamic characteristics of the study patients with indication for IHM implantation

Figure 2

Table 2. A summary of IHM sensor issues and effectiveness of interventions applied for optimising IHM performance on patients with Fontan circulation

Figure 3

Figure 2. IHM sensor discrepancies and effectiveness of interventions applied on Fontan patients.