Variation in applicator positions during interfraction high-dose rate brachytherapy in carcinoma cervix: dosimetric evaluation

Published online by Cambridge University Press:  23 April 2013

Kamlesh Passi
Affiliation:
Department of Radiation Oncology, MD Oswal Cancer Hospital, Ludhiana (Pb), India
Lalit M. Aggarwal
Affiliation:
Department of Radiotherapy & Radiation Medicine, Institute of Medical Sciences, Banaras Hindu University, Varanasi (UP), India
Rajesh Vashistha
Affiliation:
Department of Radiation Oncology, Max Superspeciality Hospital, Bathinda (Pb), India
Bikramjit Singh
Affiliation:
Department of Radiation Oncology, MD Oswal Cancer Hospital, Ludhiana (Pb), India
Parveen Kaur
Affiliation:
Department of Radiation Oncology, MD Oswal Cancer Hospital, Ludhiana (Pb), India
Sureshchandra J. Gupta
Affiliation:
Department of Physics, University of Mumbai (MH), India
Than S. Kehwar*
Affiliation:
Department of Radiation Oncology, UPMC Cancer Centers, University of Pittsburgh Cancer Institute, Pittsburgh, PA, USA
*
Correspondence to: Than S. Kehwar, Professor of Radiology, Division of Radiation Oncology, Penn State Hershey Cancer Institute, Penn State Milton S. Hershey Medical Center, 500 University Dr., CH63, Hershey, PA 17033, USA. Tel: +717-531-1523. Fax: +717-531-0446. E-mail: drkehwar@gmail.com
Rights & Permissions [Opens in a new window]

Abstract

Purpose

This study was designed to study the variations in different geometrical and dosimetric parameters.

Materials and methods

In this study, two groups comprises 21 and 28 patients, who were treated with 9·5 Gy × 2 Fx and 7·5 Gy × 3 Fx, respectively, using microselectron high-dose rate (HDR) remote control unit. All patients were analysed using orthogonal radiographs to evaluate variations in different parameters.

Results

Variations in different parameters are more in Group II patients than in Group I patients, which show that the variation in geometrical and dosimetrical parameters increases with increasing HDR number of fractions.

Conclusion

In the reporting of an outcome of multiple fractionation of HDR treatments resultant dosimetric parameters must be evaluated and must be used for clinical interpretation.

Information

Type
Original Articles
Copyright
Copyright © Cambridge University Press 2013 
Figure 0

Figure 1 (a) Anteroposterior and (b) lateral radiographs of intracavitory application.

Figure 1

Figure 2 Superimposed (a) anteroposterior and (b) lateral radiographs of intracavitory application of Group I.

Figure 2

Figure 3 Superimposed (a) anteroposterior and (b) lateral radiographs of intracavitory application of Group II.

Figure 3

Figure 4 Fusion line diagrams of (a) anteroposterior and (b) lateral radiographs of Group I of intracavitory applications.

Figure 4

Figure 5 Fusion line diagrams of (a) anteroposterior and (b) lateral radiographs of Group II of intracavitory applications.

Figure 5

Figure 6 Details of all the points and angles shown in the traced (a) anteroposterior and (b) lateral radiographs.

Figure 6

Figure 7 Illustration of fusion of the two insertions in a particular patient of Group I showing the relative position of the 9·5 Gy isodose distribution with respect to the bony land marks M1, M2 in (a) AP and (b) LAT computerised plans.

Figure 7

Figure 8 Illustration of fusion of the three insertions in a particular patient of Group II showing the relative position of the 7·5 Gy isodose distribution with respect to the bony land marks M1, M2 in (a) AP and (b) LAT computerised plans.

Figure 8

Table 1 Variations in α, β, and IUTL in Group I

Figure 9

Table 2 Variations in α, β, and IUTL in Group II

Figure 10

Table 3 Variations in distances ‘a’, ‘b’, ‘c’ and ‘d’ in Group I

Figure 11

Table 4 Variations in distances ‘a’, ‘b’, ‘c’ and ‘d’ in Group II

Figure 12

Figure 9 Variation in angle α in Group I and Group II.

Figure 13

Figure 10 Variation in angle β in Group I and Group II.

Figure 14

Figure 11 Variation in IUTL in Group I and Group II. Abbreviation: IUTL, intrauterine length.

Figure 15

Figure 12 Variation in distance ‘a’ in Group I and Group II.

Figure 16

Figure 13 Variation in distance ‘b’ in Group I and Group II.

Figure 17

Figure 14 Variation in distance ‘c’ in Group I.

Figure 18

Figure 15 Variation in distance ‘d’ in Group I and Group II.

Figure 19

Table 5 Variation in distance of RT ovoid form origin in Group I

Figure 20

Table 6 Variation in distance of RT ovoid form origin in Group II

Figure 21

Table 7 Variation in distance of LT ovoid from origin in Group I

Figure 22

Table 8 Variation in distance of LT ovoid form origin in Group II

Figure 23

Table 9 Variation in distance of external orifice of the uterus (external os) from origin in Group I

Figure 24

Table 10 Variation in distance of external orifice of the uterus (external os) form origin in Group II

Figure 25

Figure 16 Comparison of variation in distance of right ovoid from origin Group I and Group II.

Figure 26

Figure 17 Comparison of variation in distance of left ovoid from origin Group I and II.

Figure 27

Figure 18 Comparison of variation in distance of external os from origin Group I and II.

Figure 28

Table 11 Variation of point A doses in both the groups

Figure 29

Table 12 The mean volume irradiated in both the groups