Aortic migration of Amplatzer Piccolo™ ductal Occluder

Published online by Cambridge University Press:  24 February 2022

Prashant K. Minocha*
Affiliation:
Department of Pediatric Cardiology, Department of Pediatrics, Hassenfeld Children’s Hospital, NYU Langone Health, New York, NY, USA
Nicole Sutton
Affiliation:
Department of Pediatric Cardiology, Department of Pediatrics, Montefiore Medical Center, New York, NY, USA
Maya T. Crawford
Affiliation:
Department of Congenital Cardiothoracic Surgery, Hassenfeld Children’s Hospital, NYU Langone Health, New York, NY, USA
David Williams
Affiliation:
Department of Congenital Cardiothoracic Surgery, Hassenfeld Children’s Hospital, NYU Langone Health, New York, NY, USA
Michelle Ramirez
Affiliation:
Division of Pediatric Critical Care, Department of Pediatrics, Hassenfeld Children’s Hospital, NYU Langone Health, New York, NY, USA
Sujata Chakravarti
Affiliation:
Department of Pediatric Cardiology, Department of Pediatrics, Hassenfeld Children’s Hospital, NYU Langone Health, New York, NY, USA
Ralph Mosca
Affiliation:
Department of Congenital Cardiothoracic Surgery, Hassenfeld Children’s Hospital, NYU Langone Health, New York, NY, USA
T. K. Susheel Kumar
Affiliation:
Department of Congenital Cardiothoracic Surgery, Hassenfeld Children’s Hospital, NYU Langone Health, New York, NY, USA
*
Author for correspondence: P. K. Minocha, MD, Department of Pediatric Cardiology, NYU Langone Health, New York, NY 10016, USA. Tel: 212-263-3079; Fax: 212-263-8301. E-mail: Prashant.minocha@nyulangone.org
Rights & Permissions [Opens in a new window]

Abstract

We present the case of a 4-month-old, former 23-week premature baby who underwent patent ductus arteriosus device closure in the cardiac catheterisation lab with an Amplatzer Piccolo™ device at 12 weeks of life. This was complicated by late migration of the device into the aorta resulting in severe obstruction and requiring surgical intervention.

Information

Type
Brief Report
Copyright
© The Author(s), 2022. Published by Cambridge University Press
Figure 0

Figure 1.Initial cardiac catheterisation. (a) Angiogram showing a tubular patent ductus arteriosus. (b) Angiogram prior to release of device showing device release showing the device intraductully and the aortic disc to the left of the temperature probe.

Figure 1

Figure 2.Immediate post-procedure angiogram showing device placement with the pulmonary artery end in the patent ductus arteriosus and the aortic end is seen just right of the temperature probe.

Figure 2

Figure 3.(a) Immediate post-procedure echocardiogram showing intraductal discs of piccolo device and an unobstructed aortic arch by 2D and colour flow mapping. (b) A normal descending aorta doppler showing no evidence of coarctation.

Figure 3

Figure 4.Repeat cardiac catheterisation 4 weeks following initial placement. Aortic angiogram showing the device protruding into the aorta. The mid-portion of the device is past the oesophageal probe and the aortic disc is further into the aorta.

Figure 4

Figure 5.(a) The arrow points to the piccolo device which is seen at the aortic isthmus obstructing antegrade flow across the aortic arch. Colour flow mapping shows aliasing of flow in this region. (b) Continuous wave spectral doppler showing a peak systolic gradient of 54 mmHg and mean of 26 mmHg across the aortic arch.

Figure 5

Figure 6.Amplatzer Piccolo sizing chart (Year 2020).