Retrospective analysis of portal dosimetry pre-treatment quality assurance of prostate volumetric-modulated arc therapy (VMAT) plans

Published online by Cambridge University Press:  22 August 2017

Borna Maraghechi
Affiliation:
Department of Medical Physics, Grand River Regional Cancer Centre, Kitchener, ON, Canada
Jack Davis
Affiliation:
Department of Medical Physics, Grand River Regional Cancer Centre, Kitchener, ON, Canada Department of Physics and Astronomy, University of Waterloo, Waterloo, ON, Canada
Shyam Badu
Affiliation:
Department of Medical Physics, Grand River Regional Cancer Centre, Kitchener, ON, Canada
Andre Fleck
Affiliation:
Department of Medical Physics, Grand River Regional Cancer Centre, Kitchener, ON, Canada Department of Clinical Studies, Ontario Veterinary College, University of Guelph, Guelph, ON, Canada
Johnson Darko
Affiliation:
Department of Medical Physics, Grand River Regional Cancer Centre, Kitchener, ON, Canada Department of Physics and Astronomy, University of Waterloo, Waterloo, ON, Canada Department of Clinical Studies, Ontario Veterinary College, University of Guelph, Guelph, ON, Canada
Ernest Osei*
Affiliation:
Department of Medical Physics, Grand River Regional Cancer Centre, Kitchener, ON, Canada Department of Physics and Astronomy, University of Waterloo, Waterloo, ON, Canada Department of Clinical Studies, Ontario Veterinary College, University of Guelph, Guelph, ON, Canada Department of Systems Design, University of Waterloo, Waterloo, ON, Canada
*
Correspondence to: Dr Ernest Osei, Department of Medical Physics, Grand River Regional Cancer Centre, 835 King St W, Kitchener, ON, Canada, N2G 1G3. Tel: 519 749 4300, Ext: 5407. E-mail: ernest.osei@grhosp.on.ca
Rights & Permissions [Opens in a new window]

Abstract

Background

Electronic portal imaging device (EPID) offers high-resolution digital image that can be compared with a predicted portal dose image. A very common method to quantitatively compare a measured and calculated dose distribution that is routinely used for quality assurance (QA) of volumetric-modulated arc therapy (VMAT) and intensity-modulated radiation therapy treatment plans is the evaluation of the gamma index. The purpose of this work was to evaluate the gamma passing rate (%GP), maximum gamma (γmax), average gamma (γave), maximum dose difference (DDmax) and the average dose difference (DDave) for various regions of interest using Varian’s implementation of three absolute dose gamma calculation techniques of improved, local, and combined improved and local.

Methods and materials

We analyzed 232 portal dose images from 100 prostate cancer patients’ VMAT plans obtained using the Varian EPID on TrueBeam Linacs.

Results

Our data show that the %GP, γmax and γave depend on the gamma calculation method and the acceptance criteria. Higher %GP values were obtained compared with both our current institutional action level and the American Association of Physicists in Medicine Task Group 119 recommendations.

Conclusions

The results of this study can be used to establish stricter action levels for pre-treatment QA of prostate VMAT plans. A stricter 3%/3 mm improved gamma criterion with a passing rate of 97% or the 2%/2 mm improved gamma criterion with a passing rate of 95% can be achieved without additional measurements or configurations.

Information

Type
Original Articles
Copyright
© Cambridge University Press 2017 
Figure 0

Figure 1 Average gamma passing rate and 1 SD for the 3%/3 mm, 2%/2 mm and 1%/1 mm criteria for (a) improved (b) local and (c) improved+local gamma calculation methods for various regions of interest (ROIs). ROIs with low-dose thresholds of Field+0, Field+5%, Field+10%, Field+15% and MLC CIAO.

Figure 1

Table 1 Average gamma passing rate with 1 SD for improved, local and improved+local gamma calculation techniques with different acceptance criteria and various regions of interest

Figure 2

Figure 2 Distribution of average γmax with SD for 3%/3 mm, 2%/2 mm and 1%/1 mm criteria for (a) improved, (b) local and (c) improved+local gamma calculation methods within different regions of interest of Field+0, Field+5%, Field+10%, Field+15% and MLC CIAO.

Figure 3

Figure 3 Distribution of histogram representation of average γave with SD for (a) improved, (b) local and (c) improved+local gamma calculation techniques within different regions of interest (Field+0, Field+5%, Field+10%, Field+15% and MLC CIAO).

Figure 4

Table 2 The average and SD of maximum dose difference (DDmax) and average dose difference (DDave) for different low-dose thresholds