Numerical and experimental analysis of a transmission-based breast imaging system: a study of application to patients

Published online by Cambridge University Press:  07 April 2020

Elham Norouzzadeh
Affiliation:
Department of Electrical Engineering, K. N. Toosi University of Technology, Tehran, Iran
Somayyeh Chamaani*
Affiliation:
Department of Electrical Engineering, K. N. Toosi University of Technology, Tehran, Iran
Jochen Moll
Affiliation:
Department of Physics, Goethe University of Frankfurt, Frankfurt am Main 60438, Germany
Christian Kexel
Affiliation:
Department of Physics, Goethe University of Frankfurt, Frankfurt am Main 60438, Germany
Duy Hai Nguyen
Affiliation:
Department of Physics, Goethe University of Frankfurt, Frankfurt am Main 60438, Germany
Frank Hübner
Affiliation:
Institute for Diagnostic and Interventional Radiology, Goethe University Hospital Frankfurt am Main, Frankfurt am Main, Germany
Babak Bazrafshan
Affiliation:
Institute for Diagnostic and Interventional Radiology, Goethe University Hospital Frankfurt am Main, Frankfurt am Main, Germany
Thomas J. Vogl
Affiliation:
Institute for Diagnostic and Interventional Radiology, Goethe University Hospital Frankfurt am Main, Frankfurt am Main, Germany
Viktor Krozer
Affiliation:
Department of Physics, Goethe University of Frankfurt, Frankfurt am Main 60438, Germany
*
Author for correspondence: Somayyeh Chamaani, E-mail: chamaani@eetd.kntu.ac.ir
Rights & Permissions [Opens in a new window]

Abstract

Early detection of breast cancer is required to increase the chances of a successful treatment. However, current breast-imaging systems such as X-Ray mammography, breast ultrasound, and magnetic resonance imaging have technological limitations so that novel solutions are needed to address this major societal problem. The current paper considers ultra-wideband (UWB) microwave radiation in the frequency band from 1 to 9 GHz. Given by the non-ionizing nature of microwaves frequent check-ups are more feasible. In this work, we propose algorithms for qualitative and quantitative microwave breast imaging for a transmission-based UWB system. Based on numerical and experimental data, the performance of the algorithms has been investigated and compared. Finally, microwave images obtained during an initial patient study are discussed relative to corresponding X-ray images.

Information

Type
Research Paper
Copyright
Copyright © Cambridge University Press and the European Microwave Association 2020
Figure 0

Fig. 1. Schematic of the experimental breast-imaging setup.

Figure 1

Fig. 2. Illustration of the spatial sampling during one breast scan. The scanning area is limited to 50 mm × 50 mm due to the limited examination time in the clinical study of 3 min.

Figure 2

Fig. 3. Geometry of the bowtie antenna, after [44].

Figure 3

Fig. 4. Simulated reflection coefficient of a bowtie antenna, after [44].

Figure 4

Fig. 5. Simulation setup for cancer tissue detection using two bowtie antennas (Tx–Rx).

Figure 5

Fig. 6. Material properties of skin, fat, and tumor models, after [46].

Figure 6

Fig. 7. Exemplary transmission signal at antenna position 71 mm from the numerical simulation (top row) with the corresponding deviation between the pristine and the scenario with tumor (bottom row).

Figure 7

Fig. 8. RMSD of the numerical signals between the pristine and the scenario with tumor.

Figure 8

Fig. 9. Comparison between X-ray image and microwave image for patient A. The microwave image is estimated through time delays caused by the propagation through the breast when compared to air. Locations with higher relative permittivity could be interpreted as cancerous tissue. The final thickness of the compressed breast is 44 mm during microwave examination and is 41 mm during X-ray examination.

Figure 9

Fig. 10. Comparison between X-ray image and microwave image for patient B. The microwave image is formed by calculating the signal's attenuation. Locations with higher attenuation could be interpreted as cancerous tissue. The final thickness of the compressed breast is 43 mm during microwave examination and is 37 mm during X-ray examination.

Figure 10

Fig. 11. (a) Exemplary transmission signals in patient A illustrating the variations in time of arrival; (b) transmission signals obtained from patient B illustrating variations in signal attenuation.