A study of enhanced dynamic wedge dosimetry using a 2D ion chamber array detector

Published online by Cambridge University Press:  24 August 2015

S. A. Syam Kumar*
Affiliation:
Department of Radiation Oncology, Division of Radiation Physics, Malabar Cancer Centre, Thalassery, Kerala, India
P. Aparna
Affiliation:
Department of Physics, University of Calicut, Thenhippalam, Kerala, India
P. T. Anjana
Affiliation:
Department of Physics, University of Calicut, Thenhippalam, Kerala, India
C. P. Aswathi
Affiliation:
Department of Physics, University of Calicut, Thenhippalam, Kerala, India
G. P. Sitha
Affiliation:
Department of Physics, University of Calicut, Thenhippalam, Kerala, India
*
Correspondence to: Dr Surendran Nair Ambika Devi Syam Kumar, Department of Radiation Oncology, Malabar Cancer Centre, Thalasseri, Kerala, India. Tel: +91 9961443954; E-mail: skppm@yahoo.com
Rights & Permissions [Opens in a new window]

Abstract

The purpose of this work was to study the dosimetric properties of the enhanced dynamic wedge using a Seven29 ion chamber array. The PTW Seven29 ion chamber array and solid water phantoms were used for the study. Primarily, the solid water phantoms with the two-dimensional (2D) array were scanned using a computed tomography scanner at different depths. Using these scanned images, planning was performed for different wedge angles at 6 and 15 MV. A dose of 100 CGy was delivered in each case. For each delivery, the required monitoring units (MUs) were calculated. Using the same setup with a Varian Clinac iX, the calculated MU was delivered for different wedge angles. Subsequently, the different wedged dose distributions that had been obtained were analysed using Verisoft software. A shoulder-like region was observed in the profile; this region reduced as depth increased. The percentage deviation between the planned and measured doses at the shoulder region fell within the range of 0·9–4·3%. The standard deviation between planned and measured doses at shoulder region in the profile fell within 0·08±0·02 at different depths. The standard deviations between planned and measured wedge factors for different depths (2·5, 5, 10 and 15 cm) were 0·0021, 0·0007, 0·0050 and 0·0001 for 6 MV and 0·0024, 0·0191, 0·0013 and 0·0005 for 15 MV, respectively. On the basis of the studies that we performed, it can be concluded that the 2D ion chamber array is a good tool for enhanced dynamic wedge dosimetry.

Information

Type
Original Articles
Copyright
© Cambridge University Press 2015 
Figure 0

Figure 1 Setup for the scanning of solid water phantom with the 2D array. The CT scanner used is GE optima. Abbreviation: CT, computed tomography.

Figure 1

Figure 2 Verification plan in Eclipse treatment planning system using solid water phantoms with 2D array (for depth 10 cm).

Figure 2

Figure 3 Measurement setup in linac (Varian Clinac iX) where 2D array with solid water phantoms are arranged on the couch.

Figure 3

Figure 4 Beam profiles of EDW for TPS versus measured dose for 6 MV at 5 cm depth (IN and OUT orientation of 15°,45° and 60° wedges). Abbreviations: EDW, enhanced dynamic wedge; TPS, treatment planning system.

Figure 4

Figure 5 Beam profiles of EDW for TPS versus measured dose for 6 MV at 10 cm depth (IN and OUT orientation of 15°,45° and 60° wedges). Abbreviations: EDW, enhanced dynamic wedge; TPS, treatment planning system.

Figure 5

Figure 6 Beam profiles of EDW for TPS versus measured dose for 15 MV at 5 cm depth (IN and OUT orientation of 15°,45° and 60° wedges). Abbreviations: EDW, enhanced dynamic wedge; TPS, treatment planning system.

Figure 6

Figure 7 Beam profiles of EDW for TPS versus measured dose for 15 MV at 10 cm depth (IN and OUT orientation of 15°,45° and 60° wedges). Abbreviations: EDW, enhanced dynamic wedge; TPS, treatment planning system.

Figure 7

Table 1 % variation (±) between planned and measured doses at shoulder region in the profile (IN—direction)

Figure 8

Table 2 Standard deviation between doses at shoulder region in the case of eclipse planning and 2D array measurement at different depths (IN direction)

Figure 9

Table 3 Wedge factor obtained from TPS and array measurement and standard deviation between wedge factor for different wedge angles for 5 and 10 cm depth