Towards malaria microscopy at the point-of-contact: an assessment of the diagnostic performance of the Newton Nm1 microscope in Uganda

Published online by Cambridge University Press:  04 August 2014

J. RUSSELL STOTHARD*
Affiliation:
Department of Parasitology, Liverpool School of Tropical Medicine, Liverpool L3 5QA, UK
BETTY NABATTE
Affiliation:
Vector Control Division, Ministry of Health, P.O. Box 1661, Kampala, Uganda
JOSE C. SOUSA-FIGUEIREDO
Affiliation:
Department of Infectious and Tropical Diseases, London School of Hygiene and Tropical Medicine, Keppel Street, London WC1E 7HT, UK
NARCIS B. KABATEREINE
Affiliation:
Vector Control Division, Ministry of Health, P.O. Box 1661, Kampala, Uganda
*
*Corresponding author. Department of Parasitology, Liverpool School of Tropical Medicine, Liverpool L3 5QA, UK. E-mail: r.stothard@liverpool.ac.uk
Rights & Permissions [Opens in a new window]

Summary

Malaria microscopy in sub-Saharan Africa is often restricted by access to light microscopes. To address this gap, a novel portable inverted monocular microscope, the Newton Nm1, was designed and is now commercially available. Its diagnostic performance was assessed in a blinded-slide trial at ×1000 (oil) of Giemsa-stained thick blood films against a conventional microscope as undertaken by four Ugandan Ministry of Health technicians. With the Newton Nm1, diagnostic performance was: sensitivity 93·5% (95% confidence interval (CI) 78·6–99·2%), specificity 100·0% (95% CI 82·4–100·0%), positive predictive value 100·0% (95% CI 88·1–100·0%) and negative predictive value 90·5% (95% CI 69·6–98·8%). Discordance was due to a systematic error underestimating parasitaemia by ~45%; when counting Plasmodium parasites against 200 white blood cells, blood films with low parasitaemia (i.e. <100 μL−1 of blood) could be overlooked and misclassified. By contrast, specificity was excellent with no false positives encountered. Whilst proven useful, especially in resource-poor environments, it is still unclear how we can ensure the uptake of the Newton Nm1 within sub-Saharan Africa.

Information

Type
Special Issue Article
Copyright
Copyright © Cambridge University Press 2014 
Figure 0

Fig. 1. Familiarization and training of Ugandan Ministry of Health staff in the use of inverted microscopes. (A) An original metal McArthur microscope being assessed; (B) Basic training using the Newton Nm1 microscope on tripod; (C) Digital image of a Giemsa-stained thick blood film (with several malaria parasites indicated by arrows) as taken with an iPhone attached to the Newton Nm1 at ×1000 (oil) magnification.

Figure 1

Table 1. Parasitaemia within the 50 slides according to each reader and each microscope

Figure 2

Fig. 2. Bivariate scattergram demonstrating a positive correlation between Olympus CX22 and Newton Nm1 microscopes (average taken across all four readers).

Figure 3

Fig. 3. Bland–Altman difference plot comparing Newton Nm1 and Olmpus CX22 by plotting difference between microscopes the y-axis, and the mean of the two observations on the x-axis. The confidence bounds are plotted as dotted red lines, and all points within the confidence bounds are coloured green, all points outside the confidence bounds in red (sigma = 1·5). The region of agreement is the area within the confidence bounds. The blue dotted line is the ‘difference = zero line’ (optimal agreement) and the red solid line indicates the ‘average difference’.

Figure 4

Table 2. Diagnostic performance (%) of Newton Nm1 according to individual reader (and average) against ‘gold’ standard as determined by Olympus CX22

Figure 5

Table 3. Summary of the advantages and disadvantages reported by the four readers in use of the Newton Nm1 microscope