Detection of tympanic membrane movement using film patch with integrated strain gauge, assessed by optical coherence tomography: experimental study

Published online by Cambridge University Press:  27 January 2011

T Just*
Affiliation:
Department of Otorhinolaryngology, Head and Neck Surgery, University of Rostock, Germany
T Zehlicke
Affiliation:
Department of Otorhinolaryngology, Head and Neck Surgery, University of Rostock, Germany Department of Otorhinolaryngology, Military Hospital, Hamburg, Germany
O Specht
Affiliation:
Institute for Biomedical Engineering, University of Rostock, Germany
W Sass
Affiliation:
Institute of Microsystem Technology, Hamburg University of Technology, Germany
C Punke
Affiliation:
Department of Otorhinolaryngology, Head and Neck Surgery, University of Rostock, Germany
W Schmidt
Affiliation:
Institute for Biomedical Engineering, University of Rostock, Germany
E Lankenau
Affiliation:
Institute for Biomedical Optics, University of Lübeck, Germany
D Behrend
Affiliation:
Institute for Biomedical Engineering, University of Rostock, Germany
H W Pau
Affiliation:
Department of Otorhinolaryngology, Head and Neck Surgery, University of Rostock, Germany
*
Address for correspondence: Dr Tino Just, Department of Otorhinolaryngology, Head and Neck Surgery, University of Rostock, Doberaner Str 137–139, D-18057 Rostock, Germany Fax: +49 381 494 8302 E-mail: tino.just@med.uni-rostock.de
Rights & Permissions [Opens in a new window]

Abstract

Objective:

We report an ex vivo and in vivo experimental study of a device designed to measure tympanic membrane movement under normal and pathological conditions, assessed using optical coherence tomography.

Materials and methods:

We designed two types of flexible, round film patch with integrated strain gauge, to be attached to the tympanic membrane in order to measure tympanic membrane movement. Tympanic membrane attachment was assessed using optical coherence tomography. The devices were tested experimentally using an ex vivo model with varying middle-ear pressure.

Results:

Optical coherence tomography reliably assessed attachment of the film patch to the tympanic membrane, before and after middle-ear pressure changes. Strain gauge voltage changes were directly proportional to middle-ear pressure recordings, for low pressure changes. Tympanic membrane perforations smaller than 2 mm could be sealed off with the film patch.

Conclusion:

Attachment of the film patch with integrated strain gauge to the tympanic membrane was not ideal. Nevertheless, the strain gauge was able to precisely detect small pressure changes within the middle ear, in this experimental model.

Information

Type
Main Articles
Copyright
Copyright © JLO (1984) Limited 2011
Figure 0

Fig. 1 (a) The two film patch designs. The design on the right had a notch for the malleus. The design on the left includes five segmental divisions to enable a better fit with the conical shape of the tympanic membrane. (b) Film patch with integrated strain gauge (full bridge design).

Figure 1

Fig. 2 Position of the film patch with integrated strain gauge (full bridge design) against the tympanic membrane.

Figure 2

Fig. 3 Temporal bone preparation. Two plastic vein catheters were inserted into the eustachian tube opening close to the site of the internal carotid artery. Silicon gel and candle wax were used to seal the eustachian tube opening in an airtight fashion. One catheter was attached to a syringe used to create volume changes, the other catheter to a pressure sensor to measure pressure changes. The film patch with integrated strain gauge was positioned on the tympanic membrane. Data from the strain gauge and the pressure sensor were monitored simultaneously.

Figure 3

Fig. 4 (a) Optical coherence tomography scan of a healthy tympanic membrane. The three layers of the tympanic membrane are visible: from lateral to medial, the squamous, fibrous and cuboidal layers. The arrow indicates the umbo (scale bar = 200 µm). (b) Corresponding photomicrograph of the tympanic membrane (H&E; original magnification ×40).

Figure 4

Fig. 5 Optical coherence tomography scan of an atrophic tympanic membrane. The trilaminar structure of the tympanic membrane is lost (arrow) (scale bar = 100 µm).

Figure 5

Fig. 6 Optical coherence tomography scan of a chronically inflamed tympanic membrane. The three layers are not clearly detectable (scale bar = 280 µm).

Figure 6

Fig. 7 Optical coherence tomography scan of a tympanic membrane with attached film patch with integrated strain gauge, (a) before and (b) after pressure changes within the middle ear. The latter scan demonstrates a small fluid collection beneath the film patch (arrow) (scale bar = 200 µm).

Figure 7

Fig. 8 Operating microscope view of a tympanic membrane with a small perforation (arrow). A transparent film patch is positioned over the perforation. No detachment of the film patch was seen after middle-ear volume changes of less than 2 ml.

Figure 8

Fig. 9 Comparison of middle-ear pressure and strain gauge voltage over time, for low pressure changes (approximately ± 1 kPa). The strain gauge voltage is directly proportional to the pressure measurement.

Figure 9

Fig. 10 Comparison of middle-ear pressure and strain gauge voltage over time, for high pressure changes (approximately ± 3 kPa). Beyond a certain high and low pressure limit, no further change in strain gauge voltage is observed (see text for explanation).