Geometry and algorithms to expand 2θ coverage of a 2D detector

Published online by Cambridge University Press:  30 April 2018

Bob B. He*
Affiliation:
Bruker AXS Inc., Madison, Wisconsin
*
a)Author to whom correspondence should be addressed. Electronic mail: bob.he@bruker.com

Abstract

A two-dimensional (2D) diffraction pattern is an image representing the diffraction intensity distribution over the detected area. For data evaluations of various materials characterization, such as phase identification, stress, texture, and crystal size, this distribution is further converted into the intensity distribution over 2θ or γ angles. For many applications, especially phase analysis and structure refinement, it is crucial for the two-dimensional (2D) pattern to have a large 2θ range sufficient to cover as many diffraction rings as necessary. The 2θ range covered by a 2D detector is determined by the size of the detector active area and the detector distance from the sample. In order to expand the 2θ coverage with a given 2D detector, one may collect several 2D frames at various swing angles and then merge the multiple frames, or scan the 2D detector over the desired 2θ range during the data collection. This paper introduces the geometry and algorithms to produce accurate 2D diffraction patterns with expanded 2θ coverages from multiple images or scanned images.

Information

Type
Technical Article
Copyright
Copyright © International Centre for Diffraction Data 2018 
Figure 0

Figure 1. (Colour online) The angular coverage of 2D detectors of three different active areas.

Figure 1

Figure 2. (Colour online) Merge of multiple frames: (a) illustration of three detector swing angles; (b) merged 2D pattern from three flat frames.

Figure 2

Figure 3. (Colour online) Projection of multiple frames onto a cylindrical surface.

Figure 3

Figure 4. (Colour online) Projection of single frame from a flat 2D detector onto a cylindrical surface: (a) projection geometry; (b) flattened image of the projection.

Figure 4

Figure 5. (Colour online) The geometry and algorithms to project a flat 2D image onto the cylindrical surface.

Figure 5

Figure 6. (Colour online) The pixel-to-pixel projection from the flat 2D image onto the flattened cylindrical image.

Figure 6

Figure 7. (Colour online) The merge of the overlapping region: (a) detector with the rectangular active area; (b) top view of two detection planes; (c) weighting factor to combine overlapping pixels.

Figure 7

Figure 8. (Colour online) The merged image from three frames collected with Bruker Photon II™ detector from 1 µm Al2O3 powder.

Figure 8

Figure 9. (Colour online) The 2D detector scans along the detection circle while collecting diffraction signals.

Figure 9

Figure 10. (Colour online) Smearing effect when the sequential 2D frames are simply superposed from the original rectangular frames.

Figure 10

Figure 11. (Colour online) Accurate diffraction rings without smearing effect when the cylindrical projections from sequential 2D frames are combined.

Figure 11

Figure 12. (Colour online) The 2D diffraction pattern collected from Corundum by scanning a rectangular 2D detector in γ-optimized mode within 2θ range of 20–80°.