A proof of concept of a non-resonant near-field microwave microscope based on a high impedance reflectometer

Published online by Cambridge University Press:  03 June 2013

David Glay*
Affiliation:
IEMN – University of Lille 1, Av. Poincaré – CS 60069, 59652 Villeneuve d'Ascq, France. Phone: +33 (0)320197941
Adelhatif El Fellahi
Affiliation:
IEMN – University of Lille 1, Av. Poincaré – CS 60069, 59652 Villeneuve d'Ascq, France. Phone: +33 (0)320197941
Tuami Lasri
Affiliation:
IEMN – University of Lille 1, Av. Poincaré – CS 60069, 59652 Villeneuve d'Ascq, France. Phone: +33 (0)320197941
*
Corresponding author: D. Glay Email: david.glay@iemn.univ-lille1.fr
Rights & Permissions [Opens in a new window]

Abstract

In this paper, we present a non-resonant high impedance reflectometer with a reference impedance close to one of the tip probe of a near-field microwave microscope. We show that for an apex of the tip probe of 100 µm there is an optimum reference impedance close to 1 kΩ. To validate this approach a microwave circuit that makes use of lumped elements has been fabricated. A proof of concept is also explored for capacitance measurements between the tip probe and a metal plate.

Information

Type
Research Papers
Copyright
Copyright © Cambridge University Press and the European Microwave Association 2013 
Figure 0

Fig. 1. Design of the probe on Epoxy FR4-1.6 mm substrate.

Figure 1

Fig. 2. Tip-sample capacitance versus tip-conducting sample distance (HFSS simulations).

Figure 2

Fig. 3. Uncertainty of capacitance C versus reference impedance Z0.

Figure 3

Fig. 4. Bloc diagram of HIR connected to ZL.

Figure 4

Fig. 5. Schematic and signal flow graph of HIR connected to ZL.

Figure 5

Fig. 6. S21 phase-shift versus frequency for different tip-conducting sample distances.

Figure 6

Fig. 7. Equivalent impedance versus frequency of several SMD 0402 resistance values.

Figure 7

Fig. 8. Resistive matching networks: photography and experimental results.

Figure 8

Fig. 9. Power splitter with resistive matching networks: photography and experimental results.

Figure 9

Fig. 10. HMIC part of the near-field microwave microscope.

Figure 10

Fig. 11. S21 magnitude and phase-shift evolutions versus frequency.

Figure 11

Fig. 12. Measurement of S21 phase-shift evolution versus frequency for different distance d.

Figure 12

Fig. 13. Measurement of the tip probe to sample capacitance versus tip-sample distance.