Atrial septal defect closure with the new Cardia Ultrasept II™ device with interposed Goretex patch: Mexican experience – has the perforation of Ivalon’s membrane been solved?

Published online by Cambridge University Press:  27 February 2018

Mijangos-Vázquez Roberto*
Affiliation:
Pediatric Interventional Cardiology Department, Pediatric Specialties Hospital, Tuxtla Gutiérrez, Chiapas, México
García-Montes Antonio J.
Affiliation:
Pediatric Interventional Cardiology Department, National Institute of Cardiology “Ignacio Chávez”, Mexico City, Mexico
Soto-López Elena M.
Affiliation:
Aorta Clinic, Immunology Department, National Institute of Cardiology “Ignacio Chávez”, Mexico City, Mexico
Guarner-Lans Verónica
Affiliation:
Physiology Department, National Institute of Cardiology “Ignacio Chávez”, Mexico City, Mexico
Zabal Carlos
Affiliation:
Pediatric Interventional Cardiology Department, National Institute of Cardiology “Ignacio Chávez”, Mexico City, Mexico
*
Author for correspondence: M.-V. Roberto, Pediatric Interventional Cardiology Department, Pediatric Specialties Hospital, Blvd. SS Juan Pablo II y Blvd. Antonio Pariente Algarín S/N, CP. 29070, Tuxtla Gutiérrez, Chiapas. Tel: +52 961 6170700; E-mail: dr.rmijangos@gmail.com
Rights & Permissions [Opens in a new window]

Abstract

Objectives

The objective of this study was to demonstrate the safety and feasibility of using the new Cardia Ultrasept II™ device with interposed Goretex patch referring to the perforation of polyvinyl alcohol membrane.

Background

Great advances have been made in the development of devices for closure of atrial septal defect. The Cardia Ultrasept II™ with interposed Goretex patch is the modified last generation of Cardia devices, having the advantage of a super-low profile within the atria and an integral locking delivery-retrieval mechanism that ensures safe deployment. In addition, with the interposition of the Goretex, it has been possible to abolish perforation of Ivalon’s membrane as a complication.

Methods and results

Patients with ostium secundum atrial septal defect with surrounding rims with a minimum length of 5 mm and who underwent atrial septal defect closure with the new Ultrasept II™ with Goretex patch were included from two paediatric cardiac centres. Primary end point was to determine perforation of the Goretex membrane at follow-up; secondary end point included right ventricular diastolic diameter. In total, 30 patients underwent atrial septal defect closure at a median age of 6 (1–29) years. At follow-up for 6 (range, 1–15) months, freedom from perforations was 100%. A continuous decrease in right ventricular diastolic diameter was found with an initial median of 30 (25–49) mm and after catheterisation of 27.5 (18–33) mm, p=0.01, and Z-score of 2.6 (1.7–3.6) versus 1.9 (1–2.9) after procedure, p=0.01.

Conclusions

The new modified generation of the Ultrasept II™ device with interposed Goretex patch is a good alternative to achieve atrial septal defect closure safely and feasibly with no membrane perforation at follow-up.

Information

Type
Original Articles
Copyright
© Cambridge University Press 2018 
Figure 0

Figure 1 Short-axis intracardiac echocardiogram. (a) Measured basal diameter (arrow). (b) Catheter expanded diameter (tip of the arrow); the movement of the posteroinferior border can be observed (asterisk).

Figure 1

Figure 2 Final sequence once the Cardia Ultrasept II™ device has been placed. (a) Intracardiac echocardiogram. The correct placement of the device that is still being held by the cable (tip of the arrow). (b) Minnesota Manoeuvre where the correct position of both disks around the posteroinferior border can be observed (asterisk). (c) Intracardiac Echocardiogram. The device after being released (arrow). (d) Final position of the device seen by fluoroscopy.

Figure 2

Table 1 Demographic characteristics of the patients.

Figure 3

Figure 3 New Cardia Ultrasept II™ device with interposed Goretex patch.

Figure 4

Figure 4 Transthoracic echocardiogram. (a, b) Two-dimensional echo shows complete device integrity. (c, d) Color Doppler echo. No perforations observed.

Figure 5

Table 2 Changes at the end of the diastole of the right ventricle diameters before and after the closure of the atrial septal defect using the new Cardia Ultrasept II™ device with Goretex patch.