Practical lookup tables for ensuring target coverage in a clinical setup for skin cancer electron therapy

Published online by Cambridge University Press:  23 November 2017

Yongsook C. Lee
Affiliation:
Department of Radiation Oncology, University of Arizona, Tucson, AZ, USA
Yongbok Kim*
Affiliation:
Department of Radiation Oncology, University of Arizona, Tucson, AZ, USA
*
Correspondence to: Yongbok Kim, Department of Radiation Oncology, The University of Arizona, 1501N Campbell Avenue, P.O. Box 245081, Tucson, AZ 85724-5081, USA. Tel: +1 520 694 7427. Fax: +1 520 694 0228. E-mail: yongbokkim@email.arizona.edu
Rights & Permissions [Opens in a new window]

Abstract

Aim

To create practical lookup tables containing percent depth dose (PDD) and profile parameters of electron beams and to demonstrate clinical application of the lookup tables to skin cancer treatment to ensure target coverage in a clinical setup.

Materials and methods

For 6 and 9 MeV electron energies, PDDs and profiles at clinically relevant depths [i.e., R95 (distal depth of 95% maximum dose), R90, R85 and R80] were measured in water at 100 cm source-to-surface distance for an 10×10 cm2 open field and circular cutouts with diameters of 4, 5, 6, 7 and 8 cm. Then PDD parameters along with profile parameters such as width of isodose lines and penumbra at the clinically relevant depths were determined. Output factors for the cutouts were measured at dmax in water and solid water.

Results

With PDD and profile parameters, dosimetry lookup tables were generated. Based upon the lookup tables, target coverage at prescribed depths was retrospectively reviewed for three skin cancer cases. The lookup tables suggested larger cutouts for adequate target coverage.

Findings

Dosimetry lookup tables for electron beam therapy should include profile parameters at clinically relevant depths and be provided to clinicians to ensure target coverage in a clinical setup.

Information

Type
Original Article
Copyright
© Cambridge University Press 2017 
Figure 0

Table 1 Percent depth dose parameters (in centimeters) for a 10×10 cm2 open field and five circular cutouts for 6 and 9 MeV

Figure 1

Table 2 Width of isodose lines (IDLs) (in centimeters) at clinically relevant depths for 6 and 9 MeV

Figure 2

Table 3 Penumbra (in centimeters) at clinically relevant depths for 6 and 9 MeV

Figure 3

Table 4 Output factors for five circular cutouts measured at dmax in water and in solid water at 6 and 9 MeV

Figure 4

Figure 1 Squamous cell carcinoma in scalp treated with a 4-cm circular cutout using a 6 MeV electron beam.

Figure 5

Figure 2 Width of isodose lines (IDLs) and penumbra at clinically relevant depths for a 4 cm cutout for (a) 6 MeV and (b) 9 MeV. The area enclosed by an IDL is the region to receive the prescribed dose or higher.