A handheld UV-C light-emitting diode decreases environmental contamination near the operative field

Published online by Cambridge University Press:  06 January 2025

Rachael A. Turner
Affiliation:
University of Texas Rio Grande Valley School of Medicine, Edinburg, TX, USA
Roseann M. Johnson
Affiliation:
Colorado Joint Replacement, Denver, CO, USA
Yasmin Yazdani-Farsad
Affiliation:
Colorado Joint Replacement, Denver, CO, USA
Jessell Owens
Affiliation:
Colorado Joint Replacement, Denver, CO, USAGreat Basin Orthopaedics, Reno, NV, USA
Douglas A. Dennis
Affiliation:
Colorado Joint Replacement, Denver, CO, USADepartment of Mechanical and Materials Engineering, University of Denver, Denver, CO, USADepartment of Orthopaedics, University of Colorado School of Medicine, Denver, CO, USADepartment of Biomedical Engineering, University of Tennessee, Knoxville, TN, USA
Jason M. Jennings*
Affiliation:
Colorado Joint Replacement, Denver, CO, USADepartment of Mechanical and Materials Engineering, University of Denver, Denver, CO, USA
*
Corresponding author: Jason Jennings; Email: olivia.vyge@adventhealth.com

Abstract

Introduction:

Periprosthetic joint infection (PJI) may result from pathogen-to-patient transmission within the environment. High-touch surfaces (HTS) areas near the operative field from previous studies had been identified as the least likely to be thoroughly cleaned between operative cases and were utilized for this study. The purpose of this study was to assess the impact of a handheld ultraviolet-c (UV-C) light-emitting diode (LED) disinfection device on the decontamination of HTS in the operating room.

Methods:

This prospective study was conducted between 03/02/2021 and 04/20/2021. Tryptic soy agar contact plates were used to determine the bacterial load of the selected surfaces before the initiation of the case, after the case was complete, before manual cleaning, and after disinfection of the LED device. The plates were then incubated for 48 hours at 36º +/–1° C. Colony forming units (CFU) were recorded 48 hours after incubation. Mean, median, and range of CFU were recorded.

Results:

Average CFU per surface before and after the surgical case were 14.1 (range 0–200) and 13.5 (range 0–200) respectively, these were not significantly different (P = 0.9397). Manual cleaning reduced average CFU by 74% to 3.35 (range 0–200) per surface (P = 0.0162). Disinfection with the handheld LED unit further reduced the average CFU by 92% to 0.28 (range 0–4) per surface (P < 0.0001).

Conclusions:

A handheld UV-C LED disinfection device may decrease environmental contamination near the operative field in HTS areas. Further research is warranted with this technology to determine if this correlates with a decrease in PJI.

Information

Type
Original Article
Creative Commons
Creative Common License - CCCreative Common License - BY
This is an Open Access article, distributed under the terms of the Creative Commons Attribution licence (https://creativecommons.org/licenses/by/4.0/), which permits unrestricted re-use, distribution and reproduction, provided the original article is properly cited.
Copyright
© The Author(s), 2025. Published by Cambridge University Press on behalf of The Society for Healthcare Epidemiology of America
Figure 0

Figure 1.Visual location of samples collected from the anesthesia machine vitals screen.

Figure 1

Figure 2.Handheld ultraviolet light-emitting diode device used for this study.

Figure 2

Table 1.Total surface colony forming unit (CFU) counts (48 hours)

Figure 3

Table 2.Mean total colony forming unit (CFU) counts by operating room status at 48 hours