Design of a variable stiffness index finger exoskeleton

Published online by Cambridge University Press:  09 August 2021

Abhishek Attal
Affiliation:
Department of Mechanical Engineering, IIT Kanpur, Kanpur 208016, India
Ashish Dutta*
Affiliation:
Department of Mechanical Engineering, IIT Kanpur, Kanpur 208016, India
*
*Corresponding author. E-mail: adutta@iitk.ac.in
Rights & Permissions [Opens in a new window]

Abstract

This paper presents the design and experimentation of a variable stiffness index finger exoskeleton consisting of four-bar mechanisms actuated by a linear actuator. The lengths of the four-bar mechanism were optimized so that it can follow a recorded index fingertip trajectory. The mechanism has a fixed compliance at the coupler of the four-bar link and a variable compliance at the linear actuator that moves the four-bar. The skeletal shape of the coupler of the finger link has been optimized using FEM. The exoskeleton can apply a constant fingertip force irrespective of the position of the fingers.

Information

Type
Research Article
Copyright
© The Author(s), 2021. Published by Cambridge University Press
Figure 0

Figure 1. Schematic of the index/middle finger exoskeleton, the linear actuator 1 drives the four-bar mechanism and finger motion while the linear actuator 2 is used to adjust the stiffness.

Figure 1

Figure 2. Proposed finger mechanism with a linear actuator actuating the four-bar mechanism.

Figure 2

Figure 3. Basic layout of four-bar mechanism.

Figure 3

Figure 4. Three points on the desired path and corresponding diad angles.

Figure 4

Figure 5. Dynamic force analysis of the mechanism.

Figure 5

Figure 6. Surf plot showing the variation of L5, d6y, and actuator force.

Figure 6

Figure 7. Schematic of the coupler of the four-bar mechanism consisting of compliant and rigid parts.

Figure 7

Figure 8. Bezier control polygon and the corresponding curve with relation to the coupler as shown in Fig. 7.

Figure 8

Figure 9. (a) Elemental coordinate systems and (b) global coordinate system for FEM analysis.

Figure 9

Table I. Optimized parameters of the compliant coupler.

Figure 10

Figure 10. (a) shows the final shape of the coupler tip deflection with the obtained parameters and tip force of 2.5 N, while (b) shows the total deflection of the coupler and the 4-bar mechanism for a tip force of 2.5 N.

Figure 11

Figure 11. The leaf spring cantilever beam provides compliance.

Figure 12

Figure 12. Pseudo rigid body model of the cantilever.

Figure 13

Figure 13. Final CAD model of the variable compliance finger exoskeleton.

Figure 14

Figure 14. Assembled hand exoskeleton.

Figure 15

Figure 15. Assembled model of the exoskeleton showing open and close position of fingers.

Figure 16

Figure 16. Deflection of the fingertip due to compliance at two places.

Figure 17

Figure 17. Control system block diagram. The loop A controls the position of the finger during flexion and extension and the loop B controls the stiffness of the finger by changing the effective length of the cantilever beam.

Figure 18

Figure 18. Experimental setup for variable stiffness testing.

Figure 19

Table II. Variable stiffness experiment data.