Precise Limb Tourniquet Arterial Occlusion Pressure Determination using Real-Time Ultrasonography and a Capacitive-Based Force Sensor

Published online by Cambridge University Press:  18 October 2022

Jeffrey N. Wood
Affiliation:
Division of Prehospital Care, Mayo Clinic, Rochester, Minnesota, USA Department of Emergency Medicine, Mayo Clinic, Rochester, Minnesota, USA
Benjamin S. Krippendorf
Affiliation:
Department of Emergency Medicine, Mayo Clinic, Rochester, Minnesota, USA
Craig A. Blakeney
Affiliation:
Division of Prehospital Care, Mayo Clinic, Rochester, Minnesota, USA Department of Emergency Medicine, Mayo Clinic, Rochester, Minnesota, USA
Tobias Kummer
Affiliation:
Department of Emergency Medicine, Mayo Clinic, Rochester, Minnesota, USA
Alexander W. Hooke
Affiliation:
Biomechanics Research Laboratory, Mayo Clinic, Rochester, Minnesota, USA
Aidan F. Mullan
Affiliation:
Department of Health Science Research, Mayo Clinic, Rochester, Minnesota, USA
Matthew D. Sztajnkrycer*
Affiliation:
Division of Prehospital Care, Mayo Clinic, Rochester, Minnesota, USA Department of Emergency Medicine, Mayo Clinic, Rochester, Minnesota, USA
*
Correspondence: Matthew D. Sztajnkrycer, MD, PhD Mayo Clinic Department of Emergency Medicine 200 First St SW Rochester, Minnesota 55905 USA E-mail: Sztajnkrycer.matthew@mayo.edu
Rights & Permissions [Opens in a new window]

Abstract

Background:

Hemorrhage control prior to shock onset is increasingly recognized as a time-critical intervention. Although tourniquets (TQs) have been demonstrated to save lives, less is known about the physiologic parameters underlying successful TQ application beyond palpation of distal pulses. The current study directly visualized distal arterial occlusion via ultrasonography and measured associated pressure and contact force.

Methods:

Fifteen tactical officers participated as live models for the study. Arterial occlusion was performed using a standard adult blood pressure (BP) cuff and a Combat Application Tourniquet Generation 7 (CAT7) TQ, applied sequentially to the left mid-bicep. Arterial flow cessation was determined by radial artery palpation and brachial artery pulsed wave doppler ultrasound (US) evaluation. Steady state maximal generated force was measured using a thin-film force sensor.

Results:

The mean (95% CI) systolic blood pressure (SBP) required to occlude palpable distal pulse was 112.9mmHg (109-117); contact force was 23.8N [Newton] (22.0-25.6). Arterial flow was visible via US in 100% of subjects despite lack of palpable pulse. The mean (95% CI) SBP and contact force to eliminate US flow were 132mmHg (127-137) and 27.7N (25.1-30.3). The mean (95% CI) number of windlass turns to eliminate a palpable pulse was 1.3 (1.0-1.6) while 1.6 (1.2-1.9) turns were required to eliminate US flow.

Conclusions:

Loss of distal radial pulse does not indicate lack of arterial flow distal to upper extremity TQ. On average, an additional one-quarter windlass turn was required to eliminate distal flow. Blood pressure and force measurements derived in this study may provide data to guide future TQ designs and inexpensive, physiologically accurate TQ training models.

Information

Type
Original Research
Copyright
© The Author(s), 2022. Published by Cambridge University Press on behalf of the World Association for Disaster and Emergency Medicine
Figure 0

Figure 1. Ultrasound Flow Through the Brachial Artery.Note: Using a CAT7 tourniquet and a Zonare ZS3 ultrasound, arterial flow waveform was noted to be present prior to windlass revolution, absent during the phase defining occlusion, and present again after release of windlass.

Figure 1

Figure 2. Upper Extremity Blood Pressure Measurements.Note: Mean upper extremity BP, measured at the mid-bicep level, required for loss of palpable radial pulse and for loss of ultrasound-visualized brachial artery flow. Delta BP is the mean individual difference in the measured BP between loss of palpable pulse and cessation of blood flow visualized via ultrasound.Abbreviations: BP, blood pressure; SBP, systolic blood pressure.

Figure 2

Figure 3. Steady State Contact Force (N) Measurements.Note: Steady state contact force (N), as determined using a capacitative-based force sensor, for both a standard adult BP cuff and CAT7 TQ, required to eliminate both distal radial pulse by palpation and distal brachial artery flow by ultrasonography. Delta represents the mean individual difference between loss of palpable pulse and cessation of blood flow visualized via ultrasound.Abbreviations: BP, blood pressure; CAT7, Combat Application Tourniquet Generation 7; TQ, tourniquet.

Figure 3

Figure 4. Upper Extremity TQ Application.Note: Mean upper extremity windlass revolutions, measured using a CAT7 TQ at the mid-bicep level, required for loss of palpable radial pulse and for loss of ultrasound-visualized brachial artery flow. Delta represents the mean individual difference in the measured number of windlass turns between loss of palpable pulse and cessation of blood flow visualized via ultrasound.Abbreviations: CAT7, Combat Application Tourniquet Generation 7; TQ, tourniquet.

Figure 4

Figure 5. Relationship between Blood Pressure and Contact Force.Note: The association between contact force (N) and systolic blood pressure (mmHg) determined for both loss of palpable radial pulse and loss of brachial artery flow as determined by ultrasonography.Abbreviation: BP, blood pressure.