Infective endocarditis developing early after percutaneous closure of a patent ductus arteriosus in a child using the Amplatzer Duct Occluder II

Published online by Cambridge University Press:  19 March 2020

Murat M. Yilmazer*
Affiliation:
Department of Pediatric Cardiology, Izmir Dr. Behçet Uz Children’s Hospital, University of Health Sciences, İzmir35210, Turkey
Timur Meşe
Affiliation:
Department of Pediatric Cardiology, Izmir Dr. Behçet Uz Children’s Hospital, University of Health Sciences, İzmir35210, Turkey
*
Author for correspondence: M. M. Yilmazer, MD, Associate Professor, Department of Pediatric Cardiology, Izmir Dr. Behçet Uz Children’s Hospital, University of Health Sciences, İsmet Kaptan, Alsancak District, Sezer Doğan Street Number: 11, 35210 Konak/İzmir, Turkey. E-mail: drmuratmuhtar@hotmail.com
Rights & Permissions [Opens in a new window]

Abstract

A 10-month-old girl underwent patent ductus arteriosus closure with an Amplatzer Duct Occluder II. After 1 week, she was admitted to our emergency room with tachypnoea, fatigue, and fever. Consecutive blood cultures yielded vancomycin-sensitive Staphylococcus aureus. The patient was already receiving vancomycin, but the fever did not respond to this treatment. The device was successfully removed via left lateral thoracotomy.

Information

Type
Brief Report
Copyright
© The Author(s) 2020. Published by Cambridge University Press
Figure 0

Figure 1.Parasternal short-axis echocardiographic view demonstrating the Amplatzer Duct Occluder II device is more hyperechogenic than normal (in red circle).

Figure 1

Figure 2.Amplatzer Duct Occluder II device that surgically removed.