Improved visualisation of cervix applicators for magnetic resonance-only-guided brachytherapy planning

Published online by Cambridge University Press:  22 January 2014

Gary P. Liney
Affiliation:
Radiation Physics Department, Queen's Centre for Oncology and Haematology, Hull & East Yorkshire Hospitals NHS Trust, Castle Hill Hospital, Cottingham, East Yorkshire, UK
Jenny E. Marsden
Affiliation:
Radiation Physics Department, Queen's Centre for Oncology and Haematology, Hull & East Yorkshire Hospitals NHS Trust, Castle Hill Hospital, Cottingham, East Yorkshire, UK
Carl J. Horsfield
Affiliation:
Radiation Physics Department, Queen's Centre for Oncology and Haematology, Hull & East Yorkshire Hospitals NHS Trust, Castle Hill Hospital, Cottingham, East Yorkshire, UK
Tom Murray
Affiliation:
Radiation Physics Department, Queen's Centre for Oncology and Haematology, Hull & East Yorkshire Hospitals NHS Trust, Castle Hill Hospital, Cottingham, East Yorkshire, UK
David J. Manton
Affiliation:
Radiation Physics Department, Queen's Centre for Oncology and Haematology, Hull & East Yorkshire Hospitals NHS Trust, Castle Hill Hospital, Cottingham, East Yorkshire, UK
Andrew W. Beavis*
Affiliation:
Radiation Physics Department, Queen's Centre for Oncology and Haematology, Hull & East Yorkshire Hospitals NHS Trust, Castle Hill Hospital, Cottingham, East Yorkshire, UK
*
Correspondence to: Andrew W. Beavis, Radiation Physics Department, Queen's Centre for Oncology and Haematology, Castle Hill Hospital, Cottingham, East Yorkshire, HU16 5JQ, UK. Tel. 01482 461384. E-mail: andy.beavis@hey.nhs.uk.
Rights & Permissions [Opens in a new window]

Abstract

Objectives

Current guidelines for image-guided cervical cancer brachytherapy planning recommend both computed tomography (CT) and magnetic resonance imaging (MRI) for adequate visualisation of the applicator and soft tissues, respectively. MRI-only planning would be ideal as it would save time within the patient pathway and avoid the concomitant CT exposures. However, applicator visualisation on MRI is usually achieved using fluid-filled fiducial marker tubes, which can be awkward to use and suffer from unwanted air bubble artefacts. Therefore, a new fiducial-free imaging technique was developed.

Methods

A dual echo time (TE) turbo spin echo sequence was used, at 1·5 T, to provide both T2-weighted images (100 ms TE) for tissue visualisation and strongly proton density-weighted images (17 ms TE) for improved applicator visualisation. In-house software was used to automatically segment the applicator in the short TE images (using Otsu's method) and transfer the information to the long TE images to provide a single fused dataset.

Results

The method was evaluated successfully using titanium applicators in three patient cases and using a plastic applicator in a tissue-equivalent gel phantom.

Conclusions

The dual-echo technique provides a simple and efficient method for improving the visualisation of brachytherapy applicators in cervical cancer MRI images without the need for marker tubes.

Information

Type
Original Articles
Copyright
Copyright © Cambridge University Press 2014 
Figure 0

Figure 1 T2-weighted images (left-hand side) with corresponding short echo time (TE), proton density-weighted images (right-hand side) at the same slice position. The original axial images are shown on the top row, together with the reformatted coronal plane images on the bottom. These images show the difficulty in visualisation of the intrauterine applicator tube using the standard T2-weighted sequence, and how the use of a short TE improves the contrast (the tip of the applicator shown with a white arrow).

Figure 1

Figure 2 A screenshot showing how the fused applicator location (white pixels) and anatomical data images can be imported successfully into the treatment planning system in DICOM format.

Figure 2

Figure 3 (a) Axial image of the gel phantom with the intrauterine (IU) tube from the plastic applicator in situ, illustrating the difficulty in visualising the plastic applicator in a low signal (short T2) region. (b) A short echo time image of the same slice location demonstrating the IU tube clearly. (c) A repetition of the scan shown in (a), but with the marker tube inserted.