14C AMS measurement of CO2 samples with the high throughput HVE gas interface system

Published online by Cambridge University Press:  18 September 2024

G Scognamiglio*
Affiliation:
High Voltage Engineering Europa B.V., 3812 RR Amersfoort, Netherlands
M Klein
Affiliation:
High Voltage Engineering Europa B.V., 3812 RR Amersfoort, Netherlands
A Stolz
Affiliation:
Institute of Nuclear Physics, University of Cologne, 50937 Cologne, Germany
R Walet
Affiliation:
High Voltage Engineering Europa B.V., 3812 RR Amersfoort, Netherlands
D Mous
Affiliation:
High Voltage Engineering Europa B.V., 3812 RR Amersfoort, Netherlands
*
Corresponding author: G Scognamiglio; Email: info@highvolteng.com
Rights & Permissions [Opens in a new window]

Abstract

The newly designed HVE gas interface enables the AMS measurement of carbon samples in CO2 form. The CO2, e.g. resulting from the sample combustion in an elemental analyzer, is adsorbed in a zeolite trap and subsequently transferred to a motor-driven syringe. Once diluted with He, the gas mixture is transferred into the ion source of the AMS system. A carbon ion beam is formed in the ion source and mass-analyzed by the AMS system, resulting in 13C/12C and 14C/12C isotopic ratios. The HVE gas interface features two traps and two syringes to maximize the sample throughput, which results in more than 10 samples per hour. The first performance results of CO2 gas sample AMS measurements that were performed with the HVE gas interface in combination with the HVE 210 kV AMS system are presented in this paper. The measurements show that the gas interface contribution to the 14C/12C background is in the 10–15 level and to the precision is at or below 1%.

Information

Type
Conference Paper
Copyright
© The Author(s), 2024. Published by Cambridge University Press on behalf of University of Arizona
Figure 0

Figure 1 The gas interface system connects the AMS machine with different peripherals.

Figure 1

Figure 2 Schematics of the HVE gas interface. Having two zeolite traps and two syringes allows parallel operation and fast sample throughput. Switching valves direct the CO2 to the desired trap or syringe. While one trap is collecting CO2 from the EA combustion (yellow arrows), the other trap releases the CO2 to the syringe (red arrows). Similarly, when one syringe is loaded (red arrows), the other injects CO2 into the AMS ion source (green arrows). He can flow through the trap to collect the residual CO2, to the syringe to dilute pure CO2 or to the AMS ion source to purge the transfer capillary while pre-sputtering the targets (blue arrows).

Figure 2

Figure 3 The test setup used for the measurements, with the HVE gas interface accepting CO2 samples from a ThermoFisher FlashSmart EA and transferring CO2 to the HVE 210 kV AMS machine.

Figure 3

Figure 4 Left: The sample CO2 reaches the source head via the target holder and is guided onto a Ti insert placed in the target piece. Right: Section of a gas target piece.

Figure 4

Table 1 Samples used for the tests

Figure 5

Figure 5 12 C– beam current of an OxII bottle gas sample measured in the Faraday cup at the accelerator entrance. The plot shows a peak in the current due to the target insertion (A), the admission of sample CO2 into the ion source (B) and the purging of the feed capillary with He (C).

Figure 6

Figure 6 Results of the AMS measurement of 10 OxII and 10 IAEA-C3 standards. Each data point is reported with the statistical error. The orange lines indicate the calculated average value ± its standard deviation. The normalization factor, defined as the ratio between the nominal and the measured ratio, is about 1.18.

Figure 7

Table 2 AMS results of two batches of reference samples. Each batch measured 10 samples of the same kind, whose values are shown in Figure 6

Figure 8

Figure 7 Measured (non-normalized) 14C/12C ratio of different IAEA samples as a function of their nominal value.