Hostname: page-component-74656f9884-c52kr Total loading time: 0 Render date: 2026-11-09T02:41:20.507Z Has data issue: false hasContentIssue false

Estimation of monitor unit through analytical method for dynamic IMRT using control points as an effective parameter

Published online by Cambridge University Press:  02 March 2021

M. Athiyaman*
Affiliation:
Radiological Physics, Sardar Patel Medical College, Bikaner, Rajasthan, India 334003
A. Hemalatha
Affiliation:
Radiological Physics, Sardar Patel Medical College, Bikaner, Rajasthan, India 334003
Arun Chougule
Affiliation:
Radiological Physics SMS Medical College and Hospital, Jaipur, Rajasthan, India 302004
Mary Joan
Affiliation:
Radiological Physics SMS Medical College and Hospital, Jaipur, Rajasthan, India 302004
HS Kumar
Affiliation:
Radiological Physics, Sardar Patel Medical College, Bikaner, Rajasthan, India 334003
*
Author for correspondence: M. Athiyaman, Radiological Physics, Sardar Patel Medical College, Bikaner, Rajasthan, India 334003. E-mail: athiyaman.bikaner@gmail.com
Rights & Permissions [Opens in a new window]

Abstract

Introduction:

The control points (CP) play a significant role in the delivery of segmented based Intensity-Modulated Radiation Therapy (IMRT) delivery, particularly in dynamic mode. The number of segments is determined by control points and these segments will transfer from one to the other either during beam ON called dynamic delivery or during beam OFF called static delivery or step and shoot. This study was aimed at indirect estimation of the total monitor units (MU) to be delivered per field by exploiting the control points and also to find the MUs at any nth segment.

Materials and methods:

This study was performed in the Eclipse treatment planning software version 13.8.0. The details of control points, metre set weight per segment, leaf positions for each segment, field size, etc. were taken into consideration.

Results:

TPS calculated MU value and analytically estimated MU value were compared and the percentage of difference was estimated. The overall mean percentage of deviation was 1·03% between the TPS calculated method and the analytical method. The paired sample t-test was performed and, p-value <0·05, no significant difference was found. The analytical relationship determined to estimate the total number of MU delivered for any nth control point was also evaluated.

Conclusion:

The control points are a potential parameter in the conventional IMRT delivery. Through this study, we have addressed the indirect method to estimate the monitor units delivered per segment.

Information

Type
Original Article
Copyright
© The Author(s), 2021. Published by Cambridge University Press
Figure 0

Figure 1. IMRT fluence delivered in phantom and respective control points.

Figure 1

Table 1. Treatment sites have taken for MUVC verification

Figure 2

Table 2. Fields have taken for finding relationships between the control point and monitor unit

Figure 3

Figure 2. Comparison between TPS calculated MU and analytically estimated MU.

Figure 4

Figure 3. Comparison between TPS calculated dose and measured dose at I'MatriXX.

Figure 5

Table 3. Control point and comparison of MU between analytical method and TPS

Figure 6

Table 4. Comparison of measured dose and calculated dose between I’MatriXX and TPS

Figure 7

Table 5. Statistical significance

Figure 8

Figure 4: Relationship between the control point and monitor unit. (a) Relationship between CP and MU for MU/segment 1·339 and total control points 139. (b) Relationship between CP and MU for MU/segment 1·159 and total control points 201.

Figure 9

Figure 5. Relationship between the control point and monitor unit. (a) Relationship between CP and MU for MU/segment 1·064 and total control points 203. (b) Relationship between CP and MU for MU/segment 0·773 and total control points 190.