Evaluating the performance of automated external contouring tool on dose calculation of treatment planning system

Published online by Cambridge University Press:  06 February 2024

Nooshin Banaee*
Affiliation:
Medical Radiation Research Center, Central Tehran Branch, Islamic Azad University, Tehran, Iran
Mina Aeini
Affiliation:
Medical Radiation Research Center, Central Tehran Branch, Islamic Azad University, Tehran, Iran
Mehdi Salehi Barough
Affiliation:
Medical Radiation Research Center, Central Tehran Branch, Islamic Azad University, Tehran, Iran
*
Corresponding author: Nooshin Banaee; Email: nooshin_banaee@yahoo.com
Rights & Permissions [Opens in a new window]

Abstract

Introduction

The main goal of radiation therapy is to eradicate all cancer cells and minimize the damage to healthy tissues around the tumour. Treatment planning systems are used to predict the outcome of the treatment in terms of dose distribution prior to the treatment. One of the most reliable dose calculation algorithms is Monte Carlo. The aim of this study is to evaluate the performance of automated external contouring tool on dose calculation using Monte Carlo algorithm.

Materials and methods

The external contour of thorax phantom was created by automated tools of Monaco treatment planning system, and then, the IAEA-TECDOC-1583 quality assurance tests were created. Then, the treatment plans were delivered to the phantom, and the dose was measured by the Farmer ionization chamber at specific points. The external contour was corrected according to the source surface distances (SSD) which are mentioned in TECDOC-1583, and the dose was re-calculated. Finally, a comparison was made between the results.

Results

Dosimetric tests of TECDOC-1583 showed the errors ranged from −2·8% to +2·5%. In case of editing external contour and omitting fluctuations, the errors were decreased. The comparisons indicated that the most significant variation occurred in test 4 and the least changes were related to the tests 1 and 3.

Conclusions

The results of the study showed that the fluctuations of the external contour affect the calculated volume of the phantom and thus the dose. In order to obtain correct results, automated external contouring tools should be used with the correct instructions and re-checked before treatment planning.

Information

Type
Original Article
Creative Commons
Creative Common License - CCCreative Common License - BY
This is an Open Access article, distributed under the terms of the Creative Commons Attribution licence (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted re-use, distribution and reproduction, provided the original article is properly cited.
Copyright
© The Author(s), 2024. Published by Cambridge University Press
Figure 0

Table 1. The list of dosimetric tests proposed by IAEA-TECDOC-158314

Figure 1

Figure 1. (a) Adjusting the thorax phantom and (b) CT-ED phantom.

Figure 2

Figure 2. The fluctuations of external contour caused by automated tools.

Figure 3

Table 2. Results of dosimetric tests proposed by IAEA-TECDOC-1583. The external contour of the phantom is created by automated tools

Figure 4

Table 3. Results of dosimetric tests proposed by IAEA-TECDOC-1583. The fluctuations of the external contour are modified manually

Figure 5

Figure 3. CT-ED curve.