Online 13C and 14C Gas Measurements by EA-IRMS–AMS at ETH Zürich

Published online by Cambridge University Press:  17 October 2016

Cameron P McIntyre*
Affiliation:
Biogeoscience, ETH Zürich, NO, Sonnegstrasse 5, 8092, Zürich, Switzerland Laboratory of Ion Beam Physics, ETH Zürich, HPK, Otto-Stern-Weg 5, 8093 Zürich, Switzerland Current address: SUERC AMS Laboratory, University of Glasgow, Rankine Av., G75 0QF, Glasgow, UK
Lukas Wacker
Affiliation:
Laboratory of Ion Beam Physics, ETH Zürich, HPK, Otto-Stern-Weg 5, 8093 Zürich, Switzerland
Negar Haghipour
Affiliation:
Biogeoscience, ETH Zürich, NO, Sonnegstrasse 5, 8092, Zürich, Switzerland
Thomas M Blattmann
Affiliation:
Biogeoscience, ETH Zürich, NO, Sonnegstrasse 5, 8092, Zürich, Switzerland
Simon Fahrni
Affiliation:
Laboratory of Ion Beam Physics, ETH Zürich, HPK, Otto-Stern-Weg 5, 8093 Zürich, Switzerland Current address: Ionplus AG, Lerzenstrasse 12, 8953 Dietikon, Switzerland
Muhammed Usman
Affiliation:
Biogeoscience, ETH Zürich, NO, Sonnegstrasse 5, 8092, Zürich, Switzerland
Timothy I Eglinton
Affiliation:
Biogeoscience, ETH Zürich, NO, Sonnegstrasse 5, 8092, Zürich, Switzerland
Hans-Arno Synal
Affiliation:
Laboratory of Ion Beam Physics, ETH Zürich, HPK, Otto-Stern-Weg 5, 8093 Zürich, Switzerland
*
*Corresponding author. Email: cameron.mcintyre@glasgow.ac.uk.
Rights & Permissions [Opens in a new window]

Abstract

Studies using carbon isotopes to understand the global carbon cycle are critical to identify and quantify sources, sinks, and processes and how humans may impact them. 13C and 14C are routinely measured individually; however, there is a need to develop instrumentation that can perform concurrent online analyses that can generate rich data sets conveniently and efficiently. To satisfy these requirements, we coupled an elemental analyzer to a stable isotope mass spectrometer and an accelerator mass spectrometer system fitted with a gas ion source. We first tested the system with standard materials and then reanalyzed a sediment core from the Bay of Bengal that had been analyzed for 14C by conventional methods. The system was able to produce %C, 13C, and 14C data that were accurate and precise, and suitable for the purposes of our biogeochemistry group. The system was compact and convenient and is appropriate for use in a range of fields of research.

Information

Type
Advances in Physical Measurement Techniques
Copyright
© 2016 by the Arizona Board of Regents on behalf of the University of Arizona 
Figure 0

Figure 1 Schematic of the EA-IRMS–AMS system for online paired 13C and 14C gas measurements. Samples are combusted in an elemental analyzer (EA) and the product CO2 is transferred to a stable isotope ratio mass spectrometer (IRMS) for high-precision 13C measurement; 90% of the gas is split internally by the IRMS and sent to the accelerator mass spectrometer system (AMS) for 14C measurement. Details of the gas interface system and EA-IRMS can be found in Ruff et al. (2010a, 2010b).

Figure 1

Table 1 Analysis of standards by EA-IRMS–AMS. Oxalic acid and phthalic anhydride are calibration standards for both 13C and 14C.

Figure 2

Figure 2 Plot of total organic carbon (TOC) vs. depth as determined by EA-IRMS–AMS for sediment samples taken from a core from the Bay of Bengal. The relative error based on n=16 oxalic acid standards was 2%.

Figure 3

Figure 3 13C analysis by EA-IRMS–AMS of sediment samples from a core from the Bay of Bengal: (A) IRMS data with a precision of <0.1‰; (B) AMS graphite data; (C) AMS gas data using a gas ion source.

Figure 4

Figure 4 14C analysis by EA-IRMS–AMS of sediment samples from a core from the Bay of Bengal. Green triangles are conventional solid graphite data and blue circles are data from CO2 using a gas ion source. Data from 46/47 gas samples are within error of the graphite samples at the 2σ level. The single missing data point was due to a bad cathode. For comparison, red squares are conventional graphite 14C values for foraminifera from Ponton et al. (2012).

Supplementary material: PDF

McIntyre supplementary material

Table S1 and Figure S1

Download McIntyre supplementary material(PDF)
PDF 528.3 KB