Commissioning experience of the X-ray volume imaging system of an image-guided radiotherapy capable linear accelerator

Published online by Cambridge University Press:  13 January 2020

Payal Raina*
Affiliation:
Department of Radiotherapy, Rajendra Institute of Medical Sciences, Ranchi, Jharkhand, India
Susanta Kumar Sahoo
Affiliation:
Department of Radiotherapy, Rajendra Institute of Medical Sciences, Ranchi, Jharkhand, India
*
Author for correspondence: Payal Raina, Department of Radiotherapy, Rajendra Institute of Medical Sciences, E 2 Rashmi Enclave Arvind Marg, Bariatu, Bariyatu, Ranchi, Jharkhand 834009, India. E-mail: payalraina2008@gmail.com
Rights & Permissions [Opens in a new window]

Abstract

Aim:

An image-guided radiotherapy capable linear accelerator was installed at our hospital which is equipped with an X-ray volumetric imaging (XVI) system. The aim of this study was to describe the results of acceptance tests which were carried out on the XVI facility to verify the manufacturer’s specification.

Materials and methods:

The commissioning test had six elements: system safety, geometric accuracy, image quality, registration and correction accuracy, X-ray tube and generator performance, and quality assurance (QA) procedures.

Results:

We had satisfactory results for all the tests. The system passed the safety testes, and the agreement of isocentres was found to be within the tolerance limit. Imaging quality was acceptable. Registration and correction accuracy was tasted with indigenously developed phantom and positioned accurately at isocentre. X-ray tube and generator test results showed that the tube was performing properly.

Findings:

The described tests represent that the performance of the system is maintained at acceptable levels.

Information

Type
Original Article
Copyright
© The Author(s), 2020. Published by Cambridge University Press
Figure 0

Figure 1. Positioning of ball-bearing phantom.

Figure 1

Figure 2. White ball positioning.

Figure 2

Figure 3. Catphan CTP 503 phantom in position.

Figure 3

Figure 4. Uniformity module.

Figure 4

Figure 5. Contrast resolution module in Catphan CTP 503.

Figure 5

Figure 6. Spatial resolution transverse view.

Figure 6

Figure 7. Distance between two inserts.

Figure 7

Figure 8. TOR 18FG phantom with 1-mm Cu plate position.

Figure 8

Figure 9. X-ray image of TOR 18FG phantom.

Figure 9

Figure 10. Visible frequency groups.

Figure 10

Figure 11. Pelvic phantom setup.

Figure 11

Figure 12. Transverse image of automatic registration of pelvic phantom.

Figure 12

Table 1. Results of uniformity test

Figure 13

Table 2. Results of contrast insert