Combining XRD and XRF analysis in one Rietveld-like fitting

Published online by Cambridge University Press:  18 April 2017

M. Bortolotti*
Affiliation:
Dipartimento di Ingegneria Industriale, Università degli Studi di Trento, Italy
L. Lutterotti
Affiliation:
Dipartimento di Ingegneria Industriale, Università degli Studi di Trento, Italy
G. Pepponi
Affiliation:
Center for Materials and Microsystems, Fondazione Bruno Kessler, Trento, Italy
*
a) Author to whom correspondence should be addressed. Electronic mail: mauro.bortolotti@unitn.it

Abstract

X-ray diffraction (XRD) and X-ray fluorescence (XRF) are widely used analytical techniques for materials characterization; the information they provide can be considered complementary, as the former is mostly used to obtain crystallographic information and analyze phase content, whereas the latter is sensitive to elemental composition. Many researchers and technologists working in a variety of application fields already use them together in some sort of a “combined” approach, by separately performing XRD and XRF data collection and analysis on the same sample and then comparing the analytical results obtained to integrate and complement the respective analytical information. In this work, we propose a true combined approach to merge both XRD and XRF data acquisition and analysis. Custom analytical X-ray instrumentation has been developed to perform the simultaneous data acquisition, by using a single X-ray source and dedicated detectors to collect the diffracted and fluorescent X-ray photons from the same sample volume. Additionally, a combined XRD/XRF data analysis methodology has been implemented by extending Rietveld based code to incorporate the full pattern fitting of XRF spectra starting from the phases instead of a simple matrix elemental composition. We report two analytical examples from different application fields to better illustrate the capabilities of the proposed approach.

Information

Type
Technical Articles
Copyright
Copyright © International Centre for Diffraction Data 2017 
Figure 0

Figure 1. X-ray instrumentation for the combined data acquisition; an INEL Equinox 3500 diffractometer was adopted as a starting base for the development.

Figure 1

Figure 2. XRD fit for the ceramic industry waste example, showing total modeled diffraction pattern as well as the minor Calcite phase contribution (green line). Rwp: 8.5%.

Figure 2

Figure 3. Full-pattern XRF fit for the ceramic industry waste example, showing relevant element peaks. Rwp: 15.1%. The Rwp factor is not too low because the fitting is cumulative for 40 spectra at different beam incident angle. However, the fit model had very few parameters with respect to the total number of data.

Figure 3

Table I. Chemical and crystallographic quantitative analysis for the ceramic industry waste sample.

Figure 4

Figure 4. XRD fit for kaolinite sample. A small percentage of Quartz is visible. All reflections generated by the supercell Ufer single-layer model are shown. Rwp: 7.8%.

Figure 5

Figure 5. Full pattern XRF fit for kaolinite, showing relevant element peaks both from the sample and some spurious lines coming from part of the instrument such as lead shields on the collimator, brass (Cu, Zn) lines from the shutter passing through the collimator and arriving to the detector by diffraction. We take full advantage of a non-monochromatic beam to increase the number of detectable lines. With a monochromatic Mo radiation, it would have been impossible to see Zr lines. Rwp: 11.4% (40 spectra cumulative).

Figure 6

Figure 6. The 2D plot of the full XRF fit including the variation of the incident angle of the incoming beam. The incident angle was 10.5° for the first spectrum and about 49.5 for the last one. In the lower part the 40 experimental XRF spectra are reported and in the upper part the recalculated 40 ones are visible. Also, note the Cu diffraction line at 8 keV coming from the shutter as a non-continuous line.

Figure 7

Table II. Chemical and crystallographic quantitative analysis for the ceramic industry waste sample.