Hostname: page-component-74656f9884-8r8nv Total loading time: 0 Render date: 2026-11-07T02:53:40.529Z Has data issue: false hasContentIssue false

Miniaturized tunable extended eight-port hybrid coupler for 1.5 T MRI

Published online by Cambridge University Press:  03 August 2022

Zachary A. Colwell
Affiliation:
School of Biological and Health System Engineering, Arizona State University, Tempe, Arizona, USA
Sung-Min Sohn*
Affiliation:
School of Biological and Health System Engineering, Arizona State University, Tempe, Arizona, USA
*
Author for correspondence: Sung-Min Sohn, E-mail: smsohn@asu.edu
Rights & Permissions [Opens in a new window]

Abstract

This paper presents a miniaturized, tunable, high-power, eight-port hybrid coupler, based on a lumped element hybrid coupler topology. The 90° hybrid couplers are ubiquitous elements used for feeding RF coils in quadrature in magnetic resonance imaging (MRI) systems. Due to the low Larmor frequency (64 MHz) of 1.5 T MRI, distributed elements are too large for practical circuits to drive multi-port RF coils. Thus, miniaturization with MRI-compatible, non-magnetic, and high-power components is necessary. First, a miniaturized hybrid coupler is proposed for MRI systems with non-magnetic variable capacitors. Afterwards, the miniaturization methodology is applied to develop an eight-port coupler, capable of supporting both a transmit and a receive quadrature RF coil system. The high-power (up to 1 kilowatt) extended coupler measures 10 cm × 6 cm. Test results show that each port has a return loss of more than 16 dB, each input-isolated port is isolated by more than 24 dB, and each output has an insertion loss of less than 2.5 dB and output phases of 0.0°, 90.8°, 182.1°, and 278.7°.

Information

Type
Biomedical Applications
Copyright
© The Author(s), 2022. Published by Cambridge University Press in association with the European Microwave Association
Figure 0

Fig. 1. A 90° hybrid coupler ABCD transformation.

Figure 1

Fig. 2. Extended coupler schematic.

Figure 2

Table 1. Fabricated extended coupler component values

Figure 3

Fig. 3. Extended coupler physical circuit.

Figure 4

Fig. 4. Test setup with four-port vector network analyzer.

Figure 5

Fig. 5. Extended coupler simulation and bench data. (a) Measured return loss. (b) Coupling factors. (c) Phase shift.