Hostname: page-component-74656f9884-rdgjk Total loading time: 0 Render date: 2026-11-07T05:54:44.972Z Has data issue: false hasContentIssue false

Characterisation of 6MV and 10MV superficial build up dosimetry in tangential beam radiography

Published online by Cambridge University Press:  01 December 2007

Craig Lacey*
Affiliation:
Radiographer, Royal Marsden NHS Foundation Trust, Fulham Road, London,UK
Kirsti Gordon
Affiliation:
Senior Lecturer, School of Radiography, Kingston University, UK
Colin Nalder
Affiliation:
Physicist, Royal Marsden NHS Foundation Trust, Fulham Road, London, UK
*
Correspondence to: Craig Lacey, Radiographer, Royal Marsden NHS Foundation Trust, Fulham Road, UK. E-mail: craigylacey@aol.com
Rights & Permissions [Opens in a new window]

Abstract

Introduction: Although tangential radiotherapy is one of the major treatments for breast cancer, little has been done to address the skin toxicity and general dose inhomogeneity experienced in patients with larger breasts that are treated with 6MV photons. From our understanding of radiation in tissue at depth, it is proposed that 10MV photons could have a clear role in such patients through improved dose distribution. However, a greater build up depth with 10MV could mean that this energy is unacceptable.

Aims: To quantify and characterise superficial build up dosimetry in tangential breast irradiation for 6MV and 10MV photons.

Methods: Using Thermoluminescent Dosimeters (TLD’S), a comparative study was carried out investigating dose at a range of superficial depths in a phantom irradiated by tangential fields. Each delivering 2Gy for 6MV and 10MV photons.

Results: There was a 0.10Gy difference in maximum dose over a depth of 10.8 mm between 6MV and 10MV photons, along with an average difference of dose at depth of 0.09Gy.

Conclusion: Evidence has been obtained that eliminates comprise to superficial tissue if 10MV photons are used. Furthermore, reinforcement towards a more homogenous dose distribution with 10MV photons has been established.

Information

Type
Original Article
Copyright
Copyright © Cambridge University Press 2007
Figure 0

Table 1. 6MV linac output stability

Figure 1

Table 2. 10MV linac output stability

Figure 2

Figure 1. 6MV set-up and dose distribution.

Figure 3

Figure 2. 10MV set-up and dose distribution.

Figure 4

Table 3. 6MV Calibration Factor readings

Figure 5

Figure 3. Depth–dose curve for 6MV photons.

Figure 6

Figure 4. 6MV RAO and LPO depth–dose curves.

Figure 7

Table 4. 6MV: Absolute dose values at depth

Figure 8

Table 5. 10MV Calibration Factor readings

Figure 9

Table 6. 10MV: Absolute dose values at depth

Figure 10

Figure 5. Depth–dose curve for 10MV photons.

Figure 11

Figure 6. 10MV RAO and LPO depth–dose curves.

Figure 12

Figure 7. 6MV and 10MV absolute depth–dose curves.

Figure 13

Table 7. 6MV and 10MV absolute dose values and numerical difference