Commissioning a new CT simulator I: CT simulator hardware

Published online by Cambridge University Press:  01 September 2007

D. Kearns*
Affiliation:
Department of Medical Physics, Northern Ireland Cancer Centre, Belfast, N. Ireland
M. McJury
Affiliation:
Department of Medical Physics, Northern Ireland Cancer Centre, Belfast, N. Ireland
*
Correspondence to: D. Kearns, Department of Medical Physics, Northern Ireland Cancer Centre, Belfast City Hospital, Lisburn Road, Belfast BT9 7AB, N. Ireland. E-mail: denise.kearns@mpa.n-i.nhs.uk
Rights & Permissions [Opens in a new window]

Abstract

This paper reports on the commissioning tests performed on a new GE Lightspeed RT wide-bore computed tomography (CT) scanner with GE Advantage Sim software. This paper focuses on the laser marking system, CT hardware and the interfaces between each component of the system, and generalises the findings to most CT simulation systems currently available. A discussion on the commissioning of the virtual simulator software will follow in a separate paper. Three phantoms were used (two constructed in-house) to assist with a range of tests covering aspects such as the laser patient marking system, CT hardware, and connections between the CT scanner, virtual simulator system and the treatment planning system (TPS) and VARiS. Tests performed showed the CT simulation system to be working within acceptable tolerances suggested in the literature, and baseline data have been obtained against which future comparisons of system performance have been made. Where no tolerances were available, we have suggested suitable values. While considering tolerances on Hounsfield number variation that may lead to a dose error in excess of 2%, we found that in the case of low-kV CT scanning the range of Hounsfield numbers for dense bone was outside the acceptable limits for potential dose errors and so users were advised not to perform radiotherapy planning CT scans with kV below 100 kV.

Information

Type
Original Article
Copyright
Copyright © Cambridge University Press 2007
Figure 0

Figure 1. Laser check phantom and acquired 1.25 mm slice thickness CT images. (a) shows a schematic diagram of the phantom; (b) shows the central image when the phantom is aligned with lateral lasers offset in the Sup–Inf direction. When the phantom is aligned with lasers that have been correctly positioned, the three markers line up on the central slice, (c).

Figure 1

Figure 2. Schematic and central CT image of the Perspex and wire cube phantom showing the six wire markers and the two ball-bearings on either side of the phantom.

Figure 2

Table 1. Summary of tests carried out on the laser marking system

Figure 3

Table 2. Summary of CT hardware tests

Figure 4

Table 3. Acquisition protocols

Figure 5

Table 4. Hounsfield numbers and electron densities for water, lung and dense bone

Figure 6

Table 5. Hounsfield numbers and electron densities with varying kV for dense bone

Figure 7

Table 6. Summary of tests on the CT scanner-virtual simulator interface

Figure 8

Figure 3. Examples of inaccurate data transfer between the virtual simulator and the treatment planning system. (a) CT patient orientation is reversed (a) head-first to feet-first, (b) supine to prone, and (c) all structure contours are drawn on the central CT slice only.

Figure 9

Table 7. Summary of tests carried out on the virtual simulator—database interface