RF sensor dedicated to the dielectric characterization of spheroids between 500 MHz and 20 GHz

Published online by Cambridge University Press:  16 May 2023

Olivia Peytral-Rieu*
Affiliation:
LAAS-CNRS, Université de Toulouse, CNRS, UPS, Toulouse, France
David Dubuc
Affiliation:
LAAS-CNRS, Université de Toulouse, CNRS, UPS, Toulouse, France
Katia Grenier
Affiliation:
LAAS-CNRS, Université de Toulouse, CNRS, Toulouse, France
*
Corresponding author: Olivia Peytral-Rieu; Email: opeytral@laas.fr
Rights & Permissions [Opens in a new window]

Abstract

To achieve better accuracy in their investigations, biologists have recently been using three-dimensional models as intermediates between two-dimensional cell culture and the in vivo study of tissues. Some of these models are based on spheroids, which are cellular aggregates retaining many characteristics of in vivo behaviors while being still easy to use and implement in labs. To study such objects, multiple observation techniques can be used according to the objective of the study, including those using electromagnetic waves as nondestructive, noninvasive, and label-free analysis. Low-frequency (<1 MHz) ones are currently under investigation as electrochemical impedance spectroscopy. However, unlike microwaves (from 300 MHz to 60 GHz), they cannot penetrate the cell. Furthermore, most of the devices dedicated to microwave dielectric characterization are only focusing on cellular scales lower than spheroids (single cell, cell mats, or suspensions) or on tissues and organs. In this article, a microwave spectroscopy device using a coplanar waveguide adapted to the study of spheroids is presented with the dielectric characterization of fixed spheroids, with capacitance and conductance measurements from 0.5 to 20 GHz, aiming at filling the scale gap in the state of the art.

Information

Type
JNM 2022 Special Issue
Copyright
© The Author(s), 2023. Published by Cambridge University Press in association with the European Microwave Association
Figure 0

Figure 1. Experimental protocol for the manufacture of tumor spheroids.

Figure 1

Figure 2. (a) and (b) Detailed diagram of the RF device with the schematic of a spheroid at its center in pink. (a) Global view; W = signal; S = slit; G = ground. (b) Zoom in on the central part of the sensor with the dimensions of the electrodes and the capacitive gap. (c), (d), and (e) Pictures taken with an HIROX microscope. (c) Picture of the device only with the first metallization. (d) Picture of the device after the second microfabrication step: the fluidic wall. (e) Global picture of the microdevice.

Figure 2

Figure 3. Diagram of the microfabrication of the biosensor. The representation of the photosensitive resin is shown in pink,, the gold conductive layer in yellow, and the titanium layer allowing the adhesion to the quartz substrate in brown. Finally, the thickness of the SU 8 polymer is shown in green. (a) First cleaning step of the wafer. (b) Result of the deposition and exposure of the 2.5 µm NLOF resin. (c) Ti/Au metal layer (50/300 nm). (d) Results obtained after a lift-off of about 1 hour in an acetone bath. (e) Gravitational deposition of a thick layer of SU8. (f) Fluid walls obtained after exposure and development of the SU8.

Figure 3

Figure 4. (a) Presentation photo of the microwave test bench with the different elements. (b) and (c) Pictures of the device loaded with a spheroid and a polystyrene bead, respectively. (d) and (e) Capacitive contrast ΔC and conductive contrast ΔG spectra obtained with the reference fluid in blue (PBS), polystyrene beads in green and fixed spheroid in red, purple, and beige, respectively. The curves represent the average of measurements made over 25 min.