The effect of anti-scatter grids on radiation exposure during transcatheter patent ductus arteriosus closure in premature infants

Published online by Cambridge University Press:  13 December 2024

Mohamed Khallaf
Affiliation:
Division of Pediatric Cardiology, University of Iowa Stead Family Children’s Hospital, Iowa City, IA, USA Mount Sinai Kravis Children’s Heart Center, The Mount Sinai Hospital, New York, NY, USA
Osamah Aldoss
Affiliation:
Division of Pediatric Cardiology, University of Iowa Stead Family Children’s Hospital, Iowa City, IA, USA Pediatric Cardiology, Herma Heart Institute, Children’s Wisconsin and the Medical College of Wisconsin, Milwaukee, USA
Adrianne Rahde Bischoff
Affiliation:
Division of Neonatology, University of Iowa Stead Family Children’s Hospital, Iowa City, IA, USA
Jimmy Windsor
Affiliation:
Division of Pediatric Anesthesia, University of Iowa Stead Family Children’s Hospital, Iowa City, IA, USA
Joshua Stopak
Affiliation:
Division of Pediatric Anesthesia, University of Iowa Stead Family Children’s Hospital, Iowa City, IA, USA
Chaitra Mukundan
Affiliation:
Division of Pediatric Anesthesia, University of Iowa Stead Family Children’s Hospital, Iowa City, IA, USA
Patrick J. McNamara
Affiliation:
Division of Neonatology, University of Iowa Stead Family Children’s Hospital, Iowa City, IA, USA
Bassel Mohammad Nijres*
Affiliation:
Division of Pediatric Cardiology, University of Iowa Stead Family Children’s Hospital, Iowa City, IA, USA Pediatric Cardiology, King Faisal Specialist Hospital and Research Centre, Jeddah, Saudi Arabia.
*
Corresponding author: Bassel Mohammad Nijres; Email: bassel.publication@gmail.com
Rights & Permissions [Opens in a new window]

Abstract

Background:

Radiation exposure and the effect of anti-scatter grids are not well studied in premature infants during transcatheter patent ductus arteriosus closure.

This study aimed to investigate whether the use of anti-scatter grids altered the level of radiation exposure to premature infants undergoing transcatheter patent ductus arteriosus closure.

Methods and Results:

Demographic and radiation exposure data for premature infants who underwent transcatheter patent ductus arteriosus closure at the Stead Family Children’s Hospital from 10/2019 to 10/2021 were retrospectively reviewed and analyzed. Twenty-four patients (41%) underwent transcatheter patent ductus arteriosus closure with anti-scatter grids, while 34 (59%) underwent the procedure without using anti-scatter grids. At the time of catheterization, the median age, corrected age, and weight were 4.3 weeks (3.4–6), 29 weeks (28.1–30.9), and 1200g (1000-1600), respectively. Total radiation exposures for the dose area product and air kerma were 2.73 µGy.m2 (1.65–4.16), and 1.63 mGy (1.15–2.58), respectively. Radiation doses were higher in the group in whom the anti-scatter grids were utilized with dose area product of 3.33 µGy.m2 (2.39–5.43) and air kerma of 2.27 mGy (1.41–3.06) versus 1.86 µGy.m2 (1.46–3.60) and air kerma of 1.40 mGy (1.08–1.92). When radiation doses were adjusted to the radiation time, no difference in radiation exposure was noticed between the groups.

Conclusions:

Transcatheter patent ductus arteriosus closure in premature infants can be safely performed with minimal radiation exposure. In the authors’ laboratory, the use of anti-scatter grids does not impact radiation exposure in premature infants.

Information

Type
Original Article
Copyright
© The Author(s), 2024. Published by Cambridge University Press
Figure 0

Table 1. The table summarises procedure time and radiation times and doses (total and between groups). IQR = interquartile range. M–W = Mann–Whitney AK = air kerma. mGy = milliGray. DAP = dose area product. µGy.m2 = microGray meter squared

Figure 1

Figure 1. Censor diagram summarises the excluded patients. PDA = patent ductus arteriosus. IVC = inferior vena cava. B/L FV = bilateral femoral vein.

Figure 2

Figure 2. A and B show partial removal of the anti-scatter grids (white arrows) from the frontal (a) and lateral (b) camera image detectors (white arrow). C. Shows the anti-scatter grid is fully removed.