A trocar puncture robot for assisting venipuncture blood collection

Published online by Cambridge University Press:  27 March 2024

Zhikang Yang
Affiliation:
Lab of Locomotion Bioinspiration and Intelligent Robots, College of Mechanical and Electrical Engineering, Nanjing University of Aeronautics and Astronautics, Nanjing, China
Shikun Wen
Affiliation:
Lab of Locomotion Bioinspiration and Intelligent Robots, College of Mechanical and Electrical Engineering, Nanjing University of Aeronautics and Astronautics, Nanjing, China
Qian Qi
Affiliation:
Lab of Locomotion Bioinspiration and Intelligent Robots, College of Mechanical and Electrical Engineering, Nanjing University of Aeronautics and Astronautics, Nanjing, China
Zhuhai Lv*
Affiliation:
Department of Neurosurgery, Nanjing Medical University, Nanjing Brain Hospital, Nanjing, China
Aihong Ji*
Affiliation:
Lab of Locomotion Bioinspiration and Intelligent Robots, College of Mechanical and Electrical Engineering, Nanjing University of Aeronautics and Astronautics, Nanjing, China State Key Laboratory of Mechanics and Control for Aerospace Structures, Nanjing University of Aeronautics and Astronautics, Nanjing, China
*
Corresponding authors: Zhuhai Lv, Email: lvzhuhai@163.com; Aihong Ji, Email: meeahji@nuaa.edu.cn
Corresponding authors: Zhuhai Lv, Email: lvzhuhai@163.com; Aihong Ji, Email: meeahji@nuaa.edu.cn
Rights & Permissions [Opens in a new window]

Abstract

The venous blood test is a prevalent auxiliary medical diagnostic method. Venous blood collection equipment can improve blood collection’s success rate and stability, reduce the workload of medical staff, and improve the efficiency of diagnosis and treatment. This study proposed a rigid-flexible composite puncture (RFCP) strategy, based on which a small 7-degree-of-freedom (DOF) auxiliary venipuncture blood collection (VPBC) robot using a trocar needle was designed. The robot consists of a position and orientation adjustment mechanism and a RFCP end-effector, which can perform RFCP to avoid piercing the blood vessel’s lower wall during puncture. The inverse kinematics solution and validation of the robot were analyzed based on the differential evolution algorithm, after which the quintic polynomial interpolation algorithm was applied to achieve the robot trajectory planning control. Finally, the VPBC robot prototype was developed for experiments. The trajectory planning experiment verified the correctness of the inverse kinematics solution and trajectory planning, and the composite puncture blood collection experiment verified the feasibility of the RFCP strategy.

Information

Type
Research Article
Copyright
© The Author(s), 2024. Published by Cambridge University Press
Figure 0

Figure 1. Schematic diagram of the median cubital vein.

Figure 1

Figure 2. Analysis of puncture needle degrees of freedom.

Figure 2

Table I. Mechanical performance indexes.

Figure 3

Figure 3. Workflow of the venipuncture robot. (a) Rigid needle puncture blood collection process. (b) RFCP strategy.

Figure 4

Figure 4. Workflow of the venipuncture robot.

Figure 5

Figure 5. (a) 3D model of the venipuncture robot. (b) Yaw angle adjusting mechanism. (c) Pitch angle adjusting mechanism.

Figure 6

Figure 6. (a) 3D model of the puncture device. (b) Trocar needle puncture stage. (c) The needle insertion stage with flexible tube protection.

Figure 7

Figure 7. Position and orientation transformation coordinate system of the venipuncture robot.

Figure 8

Table II. MDH parameter table of venipuncture robot.

Figure 9

Figure 8. Workspace of the venipuncture robot. (a) The 3D workspace. (b) X–Y plane projection. (c) X–Z plane projection. (d) Y–Z plane projection.

Figure 10

Table III. Robot inverse solution.

Figure 11

Table IV. Verification of robot inverse solution.

Figure 12

Table V. The position and orientation of the trajectory planning fitting point.

Figure 13

Figure 9. Trajectory planning of the venipuncture robot. (a)–(c) Trajectory planning timing diagram of the venipuncture robot. (d) Variation curve of each joint. (e) The velocity curve of each joint. (f) Acceleration curve of each joint.

Figure 14

Figure 10. Prototype of the venipuncture robot. (a) The venipuncture robot. (b) The screw module and drive. (c) The motion control card. (d) The force sensor.

Figure 15

Figure 11. Path planning experiment of the venipuncture robot. (a) Initial state. (b) Position and orientation adjustment stage. (c) Puncture stage.

Figure 16

Figure 12. Venipuncture experiment on phantom arm. (a) Robot performs puncture. (b) The vacuum tube collects red liquid.