Contaminated portable medical equipment is a potential vector for transmission of healthcare-associated pathogens. Reference Donskey1 Although many devices make direct contact with patients (eg, vital signs equipment), indirect interactions are also common (ie, hands of personnel move between equipment and patients). Reference Donskey1 Whole genome sequencing in conjunction with contact tracing has provided evidence supporting transmission of pathogens such as Candidozyma auris and vancomycin-resistant enterococci (VRE) from contaminated equipment to patients. Reference Lee, White, Monahan, Jensen, Chan and van Hal2,Reference Eyre, Sheppard and Madder3 Moreover, in a recent cluster-randomized trial, manual once-daily cleaning and disinfection of shared equipment by dedicated staff resulted in improved cleaning and a significant reduction in healthcare-associated infections. Reference Browne, White and Russo4 These results suggest that improved low-level disinfection of portable equipment may reduce healthcare-associated infections.
Although it is recommended that shared equipment be cleaned and disinfected between patients, practices in real-world settings are often suboptimal. Reference Donskey1,Reference Havill, Havill, Mangione, Dumigan and Boyce5 Havill et al. Reference Havill, Havill, Mangione, Dumigan and Boyce5 reported that equipment was not being disinfected per protocol and recommended that staff receive regular education and feedback. Suwantarat et al. Reference Suwantarat, Supple, Cadnum, Sankar and Donskey6 demonstrated that portable equipment was often contaminated with pathogens. Given evidence that daily disinfection of equipment can be beneficial, Reference Browne, White and Russo4 studies are needed to evaluate current equipment cleaning and disinfection practices and levels of contamination in multiple hospitals. Here, we report findings from a culture survey of portable equipment in 31 hospitals.
Methods
The study protocol was approved by the Cleveland VA Medical Center’s institutional review board. In conjunction with a culture survey assessing Clostridioides difficile contamination after postdischarge room cleaning and disinfection, Reference Cadnum, Plumlee and Hailes7 hospitals also collected cultures from a convenience sample of shared portable devices. The study was conducted in collaboration with the Association for Professionals in Infection Control (APIC).
The 31 participating hospitals provided information on bed size and on the types of cleaning and disinfection products recommended for use on portable equipment. The sites were provided with supplies and instructions on culture collection and mailing including the types of devices to be sampled. Infection prevention personnel used Sponge Sticks with neutralizing buffer (3M, St. Paul, MN) to sample up to 10 portable devices in clinical areas that were ready for use without regard to whether it had been cleaned. For larger devices, standardized 20 × 20-cm surface areas were sampled; for smaller devices, the entire surface area was wiped with the sponges. Environmental services and clinical personnel were not made aware of the study.
The sponges were refrigerated after collection and mailed on ice to the Cleveland VA Medical Center within 1 week. Cleveland VA Medical Center personnel were blinded to the identity of the facilities. The sponges were processed for recovery of methicillin-resistant Staphylococcus aureus (MRSA), vancomycin-resistant enterococci (VRE), Candida species, and toxigenic C. difficile. Reference Cadnum, Pearlmutter and Jencson8 Microbes recovered from selective media were identified and subjected to susceptibility testing in accordance with Clinical Laboratory Standards Institute guidelines. 9 Broth enrichment cultures were processed for toxigenic C. difficile. Reference Cadnum, Plumlee and Hailes7 Fisher’s exact test was used to compare proportions of contamination among groups.
Results
Characteristics of the hospitals
The geographic location of the study hospitals is shown in the supplemental material. The average bed size was 513 (median, 350; range, 59 to 2,300 beds). Of the 31 hospitals, 23 (74.2%) reported using ready-to-use (RTU) quaternary ammonium or quaternary ammonium alcohol disinfectant wipes alone or in combination with another product for cleaning and disinfection of equipment (Supplemental material). The remaining hospitals reported primarily using dilutable quaternary ammonium products (N = 3), RTU sodium hypochlorite wipes (N = 2), dilutable peracetic acid-based products (N = 2), or RTU hydrogen peroxide wipes (N = 1).
Contamination of portable equipment
Of the 308 devices cultured, 52 (16.9%) were contaminated with 1 or more pathogens (Figure 1). Two pathogens were recovered from 10 devices and 3 were recovered from 2 devices. Candida species, including Candida albicans (N = 8) and Candida parapsilosis (N = 13), were recovered from devices in 12 hospitals and were the most common pathogens recovered (21 of 308 devices, 6.8%). VRE and toxigenic C. difficile were recovered from 19 (6.2%) and 16 (5.2%) portable devices, respectively. The supplemental material provides detailed information on the disinfectant products used and the percentages of positive cultures for the 31 individual hospitals. For the 31 hospitals, the median number of devices with contamination with 1 or more pathogens was 1 (range 0 [0%] to 7 [70%]). The percentage of contamination did not differ for hospitals using RTU wipe products versus dilutable products (45/258, 17.4% vs 7/50, 14.0%; P = .68). The percentage of C. difficile contamination did not differ for hospitals using sporicidal (N = 6) versus non-sporicidal (N = 25) disinfectants for equipment (4/60, 6.7% vs 9/248, 3.6%; P = .29).
Percentage of portable devices (N = 308) contaminated with 1 or more pathogens. MRSA, methicillin-resistant Staphylococcus aureus; VRE, vancomycin-resistant enterococci.

Figure 2 shows the percent contamination with 1 or more pathogens for different types of devices and the total numbers of each device cultured. The devices with the highest percentage of contamination included scales, isolation carts with personal protective equipment, and workstations-on-wheels.
Percentage of contamination with 1 or more pathogens for different types of devices and the total numbers of each device sampled.

Discussion
In the current study, we conducted a point-prevalence culture survey to assess contamination of in-use portable equipment in 31 hospitals. Overall, 16.9% of devices were contaminated with 1 or more pathogens. For individual hospitals, the percentage of contamination varied from 0% to 70%. Our results suggest that efforts are needed to improve cleaning and disinfection of shared portable equipment.
VRE and C. difficile were the most common bacterial pathogens recovered. Contaminated portable equipment has been implicated as a source of VRE transmission, including between different intensive care unit areas. Reference Lee, White, Monahan, Jensen, Chan and van Hal2 Most of the study hospitals reported use of non-sporicidal disinfectants for shared portable equipment. In settings with high rates of healthcare-associated CDI, use of sporicidal disinfectants for shared portable equipment and/or use of disposable equipment may be considered. Reference Cadnum, Plumlee and Hailes7 However, it is notable that the percentage of C. difficile contamination did not differ significantly for hospitals reporting use of sporicidal versus non-sporicidal disinfectants for equipment (6.7% vs 3.6%, respectively).
Contaminated surfaces have not traditionally been considered an important source for transmission of Candida spp. However, C. auris has frequently been recovered from environmental surfaces in rooms of colonized or infected patients, and an outbreak of C. auris in an intensive care unit was linked to shared temperature probes. Reference Eyre, Sheppard and Madder3 Some studies have implicated environmental surfaces, including portable equipment, as a source for acquisition of other Candida spp. For example, Sanchez et al. Reference Sanchez, Vazquez, Barth-Jones, Dembry, Sobel and Zervos10 reported recovery of a Candida parapsilosis strain from inanimate surfaces in a new intensive care unit before patients were admitted with subsequent infections due to the same strain. The ability of Candida species to survive for prolonged periods on dry and moist surfaces may increase the likelihood of transmission from the environment. Reference Eyre, Sheppard and Madder3
Our study has several limitations. The cultures were processed for only a limited number of pathogenic microbes, and therefore our results may underestimate the total burden of contamination. We did not collect information on equipment cleaning practices, and no observations or objective assessments (eg, fluorescent marker monitoring) were obtained to assess whether the devices that were cultured had been cleaned. A limited number of devices were sampled at each hospital and there is a potential selection bias because a convenience sample of devices was cultured. The study hospitals may not be representative the median bed size was 350 beds. Finally, the sampling method differed for large and small devices.
In conclusion, we found that portable medical equipment in 31 United States hospitals was frequently contaminated with healthcare-associated pathogens. Additional studies are needed to evaluate practices for cleaning and disinfection of shared equipment.
Supplementary material
The supplementary material for this article can be found at https://doi.org/10.1017/ash.2026.10816.
Acknowledgments
We thank the infection prevention personnel at the hospitals participating in the culture survey.
Financial support
This work was supported by the Clorox Company and by the US Department of Veterans Affairs Office of Research and Development as part of funding for VASeqCURE, which in turn received funding from the American Rescue Plan Act funds. Clorox Company personnel provided input on study design but did contribute to data analysis or writing, editing, or approval of the manuscript.