Measurement of the photon and thermal neutron doses of contralateral breast surface in breast cancer radiotherapy

Published online by Cambridge University Press:  05 August 2019

Hamed Bagheri
Affiliation:
AJA Radiation Sciences Research Center, AJA University of Medical Sciences, Tehran, Iran
Razzagh Abedi Firouzjah
Affiliation:
Medical Physics Department, Faculty of Medicine, Babol University of Medical Sciences, Babol, Iran
Bagher Farhood*
Affiliation:
Department of Medical Physics and Radiology, Faculty of Paramedical Sciences, Kashan University of Medical Sciences, Kashan, Iran
*
Author for correspondence: Bagher Farhood, Department of Radiology and Medical Physics, Faculty of Paramedical Sciences, Kashan University of Medical Sciences, Kashan, Iran. Tel: +98 9129234207. Fax: +98 31 55548883. E-mail: bffarhood@gmail.com
Rights & Permissions [Opens in a new window]

Abstract

Introduction and purpose:

During the radiation therapy of tumoral breast, the contralateral breast (CB) will receive scattered doses. In the present study, the photon and thermal neutron dose values received by CB surface during breast cancer radiation therapy were measured.

Materials and methods:

The right breast region of RANDO phantom was considered as CB, and the measurements of photon and thermal neutron dose values were carried out on this region surface. The phantom was irradiated with 18 MV photon beams, and the dose values were measured with thermoluminescent dosimeter (TLD-600 and TLD-700) chips for 11 × 13, 11 × 17 and 11 × 21 cm2 field sizes in the presence of physical and dynamic wedges.

Results:

The total dose values (photon + thermal neutron) received by the CB surface in the presence of physical wedge were 12·06%, 15·75% and 33·40% of the prescribed dose, respectively, for 11 × 13, 11 × 17 and 11 × 21 cm2 field sizes. The corresponding dose values for dynamic wedge were 9·18%, 12·92% and 29·26% of the prescribed dose, respectively. Moreover, the results showed that treatment field size and wedge type affect the received photon and thermal neutron doses at CB surface.

Conclusion:

According to our results, the total dose values received at CB surface during breast cancer radiotherapy with high-energy photon beams are remarkable. In addition, the dose values received at CB surface when using a physical wedge were greater than when using a dynamic wedge, especially for medial tangential fields.

Information

Type
Original Article
Copyright
© Cambridge University Press 2019
Figure 0

Figure 1. Tangential fields of breast radiation therapy and placement of TLD chips on the RANDO phantom.

Figure 1

Table 1. Photon and thermal neutron dose values received at contralateral breast (CB) surface relating to wedge types and different treatment field sizes

Figure 2

Figure 2. Photon and neutron dose values received at contralateral breast (CB) surface for different treatment field sizes with physical (a) and dynamic (b) wedges.

Figure 3

Figure 3. Effect of wedge types on the received photon (a) and neutron (b) dose values at contralateral breast (CB) surface for different field sizes.

Figure 4

Figure 4. Effect of treatment field size and wedge type on photon and neutron dose values received at the contralateral breast (CB). (a), (b) and (c) relate to 11 × 13, 11 × 17 and 11 × 21 cm2 treatment filed sizes, respectively.

Figure 5

Table 2. Comparison of dose values received at contralateral breast (CB) surface and CB relating to wedge types and different treatment filed sizes

Figure 6

Figure 5. Comparison of dose values received at contralateral breast (CB) surface and CB in the presence of physical (a) and dynamic (b) wedges for different treatment field sizes.

Figure 7

Figure 6. Comparison of dose values received at contralateral breast (CB) surface and CB in presence of wedge type for 11 × 13 (a), 11 × 17 (b) and 11 × 21 (c) cm2 field sizes.