Semi-automated speckle-tracking for quantitative right ventricular assessment in children and adolescents

Published online by Cambridge University Press:  15 August 2019

Aron K. McCloud*
Affiliation:
Division of Pediatric Cardiology, Advocate Children’s Hospital, Oak Lawn, IL, USA Division of Pediatric Cardiology, Jacobs School of Medicine and Biomedical Sciences, University at Buffalo, Buffalo, NY, USA
Joanna Lowisz
Affiliation:
Division of Pediatric Cardiology, Advocate Children’s Hospital, Oak Lawn, IL, USA
David A. Roberson
Affiliation:
Division of Pediatric Cardiology, Advocate Children’s Hospital, Oak Lawn, IL, USA
Cheryl A. Lefaiver
Affiliation:
Center for Pediatric Research, Advocate Children’s Hospital, Oak Lawn, IL, USA
Jamie S. Penk
Affiliation:
Division of Pediatric Cardiology, Advocate Children’s Hospital, Oak Lawn, IL, USA
*
Author for correspondence: A. K. McCloud, DO, Division of Pediatric Cardiology, University of Virginia, 629 Cedar Creek Grade, Suite B, Winchester, VA 22601, USA. Tel: (540) 678-3950; Fax: (540) 678-3954; E-mail: akm6v@virginia.edu
Rights & Permissions [Opens in a new window]

Abstract

Background:

Assessment of right ventricular size and function is an important part of the clinical cardiac evaluation; however, these quantitative measures are challenging by echocardiography. Automated software could be useful in place of manual measurements and qualitative assessment. This study evaluates a semi-automated software by comparing automated to manual measures in normal children.

Methods:

Neonates to adolescents with normal echocardiograms were prospectively enrolled. Measurements were performed using manual techniques and semi-automated software (EchoInsight®, Epsilon Imaging, Ann Arbor, Michigan, United States of America). Right ventricular measurements included end-diastolic and end-systolic area, fractional area change, chamber dimensions, and tricuspid annular plane systolic excursion. Agreement between manual and semi-automated measures was compared.

Results:

Echocardiograms for 233 patients were included in the analysis. Intra- and inter-observer reliabilities for semi-automated measures were good with intraclass correlation coefficients all over 0.9 and 0.85, respectively. There was very strong correlation between manual and semi-automated methods for areas and dimensions (r = 0.93–0.99) and low bias (1.4–10.8%). For functional measures, tricuspid annular plane systolic excursion measures correlated well (r = 0.84), but fractional area change did not (r = 0.50). Both demonstrated significant bias (33.5–43.0%). The semi-automated method consistently underestimated fractional area change with a mean of 26.6% versus a manual mean of 36.1%.

Conclusions:

The semi-automated software is capable of generating quantitative right ventricular measures in children with good reliability. The software demonstrates very good correlation and low bias when compared to manual methods for right ventricular areas and dimensions. There is a significant difference between manual and semi-automated techniques for the functional measures.

Information

Type
Original Article
Copyright
© Cambridge University Press 2019 
Figure 0

Figure 1. Apical four-chamber view with EchoInsight® tracking of the right ventricular myocardium in end-diastole. This tracking is used to generate tracings of dimensions. The arrow represents the tracking of the direction and displacement of the lateral tricuspid valve hinge used to measure tricuspid annular plane systolic excursion.

Figure 1

Figure 2. Apical four-chamber view with EchoInsight® tracking of the right ventricular myocardium in end-systole.

Figure 2

Figure 3. EchoInsight® tracing of right ventricular end-diastolic area.

Figure 3

Figure 4. EchoInsight® tracing of right ventricular end-systolic area.

Figure 4

Figure 5. EchoInsight® tracing of right ventricular basilar, mid-cavitary, and longitudial dimensions.

Figure 5

Table 1. Demographics and basic echo information

Figure 6

Table 2. Intra and inter-observer reliability

Figure 7

Table 3. Manual and semi-automated analysis

Figure 8

Table 4. Manual and semi-automated bias/correlation

Figure 9

Table 5. Relationship between age and percentage of mean difference

Figure 10

Figure 6. Scatterplots of percentage of mean difference versus age. BASL DIAM=basal diameter; EDA=end-diastolic area; ESA=end-systolic area; FAC=fractional area change; LONG DIMEN=length; MIDCAV DIAM=mid-cavitary diameter; RV=right ventricle; TAPSE=tricuspid annular plane systolic excursion.

McCloud et al. supplementary material

McCloud et al. supplementary material 1

Download McCloud et al. supplementary material(Video)
Video 32.1 MB