Conventionally Heated Microfurnace for the Graphitization of Microgram-Sized Carbon Samples

Published online by Cambridge University Press:  08 November 2016

Bin Yang*
Affiliation:
Australian Nuclear Science and Technology Organisation, New Illawarra Road, Lucas Heights, NSW 2234, Australia
A M Smith
Affiliation:
Australian Nuclear Science and Technology Organisation, New Illawarra Road, Lucas Heights, NSW 2234, Australia
*
* Corresponding author. Email: byz@ansto.gov.au.
Rights & Permissions [Opens in a new window]

Abstract

A new type of miniaturized, externally heated graphitization reaction furnace, the microconventional furnace (MCF), was constructed following our development of the laser heated furnace (LHF). The MCF is comprised of a gas reactor, a cold finger cooling system, and a compact resistive heater, which can raise the temperature of the hot finger to 850°C. The gas reactor is provided with three integrated valves to connect with the hydrogen/vacuum manifold, to isolate the reactor, and to connect with sample vessels. We made two types of MCF: the type 1 furnace (volume of 0.9 mL), with an integral stainless steel cold finger, and the type 2 furnace (volume from 1.3 to 10 mL), with a changeable glass cold finger. The MCF is designed for above atmospheric pressure (up to 2500 mbar) operation to decrease the overall graphitization time and improve the carbon yield. The MCF provides an effective solution for producing graphite from carbon dioxide (CO2) sample gas from 5 to 2000 µg of carbon with only 0.083 μg of 100 pMC extraneous carbon added. Cross-contamination tests show that the MCFs have no memory effect from previous samples.

Information

Type
Chemical Pretreatment Approaches
Copyright
© 2016 by the Arizona Board of Regents on behalf of the University of Arizona 
Figure 0

Figure 1 Block diagram of the MCF showing three main components: gas reactor, tube furnace, and cold finger cooling system (separated by dashed lines). The light blue (gray) regions represent the vacuum manifold with pressure gauges and valves. Also provided here the part name and company of each electronic unit.

Figure 1

Figure 2 Photograph of a type 2 MCF with the following parts: (1) PID temperature controller of the tube heater; (2) tube heater; (3) gas reactor with three valves; (4) a spare sample vessel contacting with spare port; (5) a changeable glass cold finger; (6) sample vessel fitting in the sample port; (7) a cold finger cooling system for trapping both water and CO2 gas; and (8) temperature controller for the cold finger.

Figure 2

Figure 3 (a) A vertical cross-section through central of the core of type 1 MCF showing the reactor with the cold finger, 3.5-mm-diameter pressure sensor, reactor valve, and the port for the glass tube. (b) A horizontal cross-section through the core shows the interconnections between the valves and the tubes.

Figure 3

Figure 4 Photograph of an opened tube heater: (1) stainless steel cover; (2) bottom plate; (3) thermocouple wire; (4) stainless steel foil; (5) quartz wool; and (6) central heating tube. Items (1), (2), (4), and (5) provide very good thermal isolation of the heater.

Figure 4

Figure 5 The vacuum manifold of MCF and laser heated furnaces: (1) LHFa; (2) LHFb; (3)MCF C1; (4) C2; (5) C3; (6) C4; (7) H2 valves; (8) MKS Baratron gauge; (9) Pfeiffer PKR 251 gauge; (10) main vacuum valve; (11) turbomolecular pumping station; (12) system cold finger; (13) 90° angle valve link to graphitization units; and (14) vacuum manifold.

Figure 5

Figure 6 Pressure versus time recorded during the graphitization of 5- and 10-µg C samples. The 10-µg sample plot shows a variation of pressure during graphitization: after 2 min, the temperature of the cold finger (TCF) was raised from –160°C to –70°C so that the CO2 was released from the cold finger; after 5 min, the pressure PCO2 was recorded, then TCF was again set to –160°C so the pressure dropped to zero; after 13 min, H2 was fed into the reactor, the isolation valve closed, and the cold finger set to –70°C; after 15 min, PCO2+H2 was recorded and the hot finger temperature set to 600°C; the pressure increased due to thermal expansion up to a peak at 16 min followed by a rapid decrease during graphitization; the reaction was complete after 29 min when the heater was pushed back to cool the hot finger to RT; after 32 min, the final pressure PEND was recorded and the manifold valve was opened, dropping the pressure to zero.

Figure 6

Figure 7 Pressure recorded during the graphitization of 113- and 980-µg C samples.

Figure 7

Figure 8 The efficiencies of carbon yield for graphitization of standard and blank samples with different masses.

Figure 8

Table 1 Carbon mass, 13C current, mass of Fe catalyst (reduced Fe2O3), and current/Cmass for MCF C1 and C2. Each sample was prepared by graphitizing 14C-depleted CO2 gas. The currents were normalized to a large natural graphite target that produced mean 13C4+ current of 606 nA in the first part of the run and 660 nA after the cesium flow was increased. The listed currents correspond to the low cesium flow.

Figure 9

Figure 9 Original data of a 31.8-µg C standard CO2 sample graphitized by C1 and measured by AMS using ANTARES. This sample was measured for two non-concurrent 10-min periods. Diamonds show the 14C/13C ratio, while dots show the 14C/13C current. Solid lines are the mean 14C/13C ratios for each 10-min period.

Figure 10

Figure 10 The extraneous carbon added during graphitization by introducing 14C-depleted CO2 sample of varying mass and plotting the measured activity versus 1/mass, was fitted by the equation: Y(pMC)=0.03+a * X (1/μg C): for C1 and C2 and for C3 and C4.

Figure 11

Figure 11 The SSV has two valves. It is a very important tool at ANSTO for preparing small standard samples.

Figure 12

Table 2 Experimental parameters of preparation of the small OX2 samples together with the pMC values by AMS.

Figure 13

Table 3 The memory was tested in the MCF C1, C2, C3, and C4 by preparing a ~30-µg C blank (Blank pre-OX2), followed by a ~120-µgC OX2, followed again by another ~30-µg C blank (Blank Post-OX2). The result indicates that MCFs have no memory of previous samples. The pMC values of Blank Pre-OX2 are higher than the pMC values of Blank Post-OX2 due to the contamination of the transfer line during the transfer of the first set of the four CO2 samples. In addition, comparing with Table 2, it can be noted that the lower pMC of OX2 samples because the CO2 gas was obtained from directly combusted oxalic acid without further splitting by SSV to reduce the blank Cex during the combustion.