Filtration evaluation and clinical use of expired elastomeric P-100 filter cartridges during the COVID-19 pandemic

Published online by Cambridge University Press:  27 May 2020

Harsh H. Patolia
Affiliation:
Virginia Tech Carilion School of Medicine, Roanoke, Virginia
Jin Pan
Affiliation:
Civil and Environmental Engineering, Virginia Tech, Blacksburg, Virginia
Charbel Harb
Affiliation:
Civil and Environmental Engineering, Virginia Tech, Blacksburg, Virginia
Linsey C. Marr
Affiliation:
Civil and Environmental Engineering, Virginia Tech, Blacksburg, Virginia
Anthony W. Baffoe-Bonnie*
Affiliation:
Virginia Tech Carilion School of Medicine, Roanoke, Virginia Carilion Clinic Roanoke Memorial Hospital, Roanoke, Virginia
*
Author for correspondence: Anthony W. Baffoe-Bonnie, E-mail: abaffoe6@vt.edu
Rights & Permissions [Opens in a new window]

Abstract

Information

Type
Letter to the Editor
Creative Commons
Creative Common License - CCCreative Common License - BY
This is an Open Access article, distributed under the terms of the Creative Commons Attribution licence (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted re-use, distribution, and reproduction in any medium, provided the original work is properly cited.
Copyright
© 2020 by The Society for Healthcare Epidemiology of America. All rights reserved.
Figure 0

Fig. 1. Filtration efficiency of the P100 filter cartridge and 3M 1860 (N95) for salt (NaCl) particles 0.04–1.0 μm in aerodynamic diameter. Error bars represent standard deviations across 2 cartridges, each tested twice.