Initial assessment of a novel delivery system (NuDEL™ ®) for the covered Cheatham-Platinum stent

Published online by Cambridge University Press:  15 June 2017

Gareth J. Morgan*
Affiliation:
Departments of Paediatric and Adult Congenital Cardiology, Evelina London Children’s Hospital, Guy’s and St Thomas’ NHS Foundation Trust, King’s College London, London, United Kingdom
Damien Kenny
Affiliation:
Department of Paediatric Cardiology, Our Lady’s Hospital for Sick Children, Crumlin, Dublin, Ireland
Christopher Duke
Affiliation:
East Midlands Congenital Heart Centre, Glenfield Hospital, Leicester, United Kingdom
Kevin P. Walsh
Affiliation:
Department of Paediatric Cardiology, Our Lady’s Hospital for Sick Children, Crumlin, Dublin, Ireland
Shakeel A. Qureshi
Affiliation:
Departments of Paediatric and Adult Congenital Cardiology, Evelina London Children’s Hospital, Guy’s and St Thomas’ NHS Foundation Trust, King’s College London, London, United Kingdom
*
Correspondence to: Dr G. J. Morgan, Department of Congenital Cardiology, Children’s Hospital of Colorado, University of Colorado, Denver, CO 80045, United States of America. Tel: +1 720 777 3379; Fax: +1 720 777 7177; E-mail: drgarethjmorgan@gmail.com
Rights & Permissions [Opens in a new window]

Abstract

Aims

We sought to evaluate the first-in-man use of a new system for implantation of covered stents in patients with complex structural and CHD.

Methods and results

Retrospective data were collected of the first 13 NuDEL™ delivery systems used in patients. The NuDEL™ comprises a covered Cheatham-Platinum stent mounted on a balloon-in-balloon and pre-loaded in a long delivery sheath. Data were collected from three centres in the United Kingdom and Ireland. A total of 13 covered stents were delivered via 12 NuDEL™ delivery systems in 12 patients. Among them, six patients had coarctation of the aorta, five patients had right ventricular outflow tract stenosis, and one patient had severe stenosis of a Mustard systemic venous baffle. There were no complications, and all the stents were deployed in the desired position with satisfactory haemodynamic results.

Conclusions

The development of a bespoke system of a pre-mounted, pre-loaded covered stent may negate some of the technical challenges that complicate large-calibre stent deployment. Our preliminary results suggest that the NuDEL™ system is a safe and effective means of covered stent deployment in challenging anatomy.

Information

Type
Original Articles
Copyright
© Cambridge University Press 2017 
Figure 0

Figure 1 The NuDEL™ system. The stent and balloon complex are exposed by pulling the sheath back over the shaft of the balloon. Inset (a) illustrates the system fully covered; (b) illustrates the sheath partially withdrawn over the stent; and the main picture (c) shows the stent and balloon fully uncovered and ready to deploy.

Figure 1

Figure 2 NuDEL™ being used in challenging anatomy. (a) Atretic aortic coarctation, which after being crossed with a wire (b) was then crossed without pre-dilation with the NuDEL™ system (c). (d and e) Highly angulated and calcified complex right ventricular outflow (RVOT). This was traversed easily with the NuDEL™ system (f) to allow placement of a covered Cheatham-Platinum in the RVOT before placing a percutaneous pulmonary valve.

Figure 2

Table 1 Patient demographics and patient details.