Hostname: page-component-74656f9884-g96k6 Total loading time: 0 Render date: 2026-11-10T10:30:16.270Z Has data issue: false hasContentIssue false

Early dehiscence of a tricuspid valve annuloplasty ring in an adolescent with hypoplastic left heart syndrome presenting with unconjugated hyperbilirubinemia

Published online by Cambridge University Press:  22 November 2024

David Kielmayer*
Affiliation:
Department of Pediatric Cardiology, Kepler University Hospital, Linz, Austria
Eva Sames-Dolzer
Affiliation:
Division of Pediatric and Congenital Heart Surgery, Kepler University Hospital, Linz, Austria Medical Faculty, Johannes Kepler University Linz, Linz, Austria
Gerald Tulzer
Affiliation:
Department of Pediatric Cardiology, Kepler University Hospital, Linz, Austria Medical Faculty, Johannes Kepler University Linz, Linz, Austria
*
Corresponding author: David Kielmayer; Email: david.kielmayer@kepleruniklinikum.at
Rights & Permissions [Opens in a new window]

Abstract

We report a unique case of an adolescent patient with Fontan physiology presenting with unconjugated hyperbilirubinemia due to dehiscence of a tricuspid valve annuloplasty ring.

Information

Type
Brief Report
Copyright
© The Author(s), 2024. Published by Cambridge University Press
Figure 0

Figure 1. (a) Two-dimensional transthoracic echocardiography apical view indicating dehiscence of the tricuspid annuloplasty ring from the septal valve aspect during diastole. (b) Colour Doppler image during diastole with blood inflow separated by the septal part of the ring and (c) tricuspid regurgitation during systole. F = extracardiac Fontan connection, LA = left atrium, RA = right atrium, RV = right ventricle, TVAL = anterior leaflet of tricuspid valve, *septal leaflet of tricuspid valve, empty arrow heads: area of dehiscence, filled arrows: tricuspid annuloplasty ring.

Figure 1

Figure 2. Intraoperative surgeons view showing the loose part of the annuloplasty ring held by the forceps.

Figure 2

Figure 3. Dynamics of total serum (blue line) and unconjugated (green line) bilirubin (mg/dL) from readmission showing dramatic drop of these parameters immediately after implantation of the mechanical tricuspid valve on day 21 indicated by the red vertical line. Follow-up total serum bilirubin consistently showed normal values until day 70 after readmission; unconjugated bilirubin was too low to be determined.