Calibration of X-ray imaging devices for accurate intensity measurement

Published online by Cambridge University Press:  15 June 2012

Michael J. Haugh*
Affiliation:
National Security Technologies, LLC, 161 S. Vasco Rd. Suite A, Livermore, California 94550
Michael R. Charest
Affiliation:
National Security Technologies, LLC, 161 S. Vasco Rd. Suite A, Livermore, California 94550
Patrick W. Ross
Affiliation:
National Security Technologies, LLC, 161 S. Vasco Rd. Suite A, Livermore, California 94550
Joshua J. Lee
Affiliation:
National Security Technologies, LLC, 161 S. Vasco Rd. Suite A, Livermore, California 94550
Marilyn B. Schneider
Affiliation:
Lawrence Livermore National Laboratory, 7000 East Ave., Livermore, California 94550
Nathan E. Palmer
Affiliation:
Lawrence Livermore National Laboratory, 7000 East Ave., Livermore, California 94550
Alan T. Teruya
Affiliation:
Lawrence Livermore National Laboratory, 7000 East Ave., Livermore, California 94550
*
a) Author to whom correspondence should be addressed. Electronic mail: haughmj@nv.doe.gov

Abstract

National Security Technologies (NSTec) has developed calibration procedures for X-ray imaging systems. The X-ray sources that are used for calibration are both diode type and diode/fluorescer combinations. Calibrating the X-ray detectors is a key to accurate calibration of the X-ray sources. Both energy dispersive detectors and photodiodes measuring total flux were used. We have developed calibration techniques for the detectors using radioactive sources that are traceable to the National Institute of Standards and Technology (NIST). The German synchrotron at Physikalische Technische Bundestalt (PTB) was used to calibrate the silicon photodiodes over the energy range from 50 to 60 keV. The measurements on X-ray cameras made using the NSTec X-ray sources included quantum efficiency averaged over all pixels, camera counts per photon per pixel, and response variation across the sensor. The instrumentation required to accomplish the calibrations is described. The X-ray energies ranged from 720 to 22.7 keV. The X-ray sources produce narrow energy bands, allowing us to determine the properties as a function of X-ray energy. The calibrations were done for several types of imaging devices. There were back and front illuminated CCD (charge-coupled device) sensors, and a CID (charge injection device) type camera. The CCD and CID camera types differ significantly in some of their properties that affect the accuracy of the X-ray intensity measurements. All the cameras discussed here are silicon based. The measurements of the quantum efficiency variation with the X-ray energy are compared to the models for the sensor structure. The cameras that are not back-thinned are compared to those that are.

Information

Type
Technical Articles
Copyright
Copyright © International Centre for Diffraction Data 2012
Figure 0

Figure 1. (Color online) The HEX system layout and the fluorescer source arrangement.

Figure 1

Figure 2. (Color online) A typical spectrum from the HEX source observed using a solid state Ge energy dispersive detector.

Figure 2

Figure 3. (Color online) Schematic diagram of the MRXS comprising three separate vacuum chambers. The multiple diode source is in the center chamber.

Figure 3

Figure 4. (Color online) The X-ray spectrum obtained using a Ti anode and a Ti filter to produce an X-ray beam comprising mainly the Ti Kα and Kβ spectral lines. It is observed using a Si solid state energy dispersive detector.

Figure 4

TABLE I. Properties of semiconductors discussed in this section.

Figure 5

TABLE II. Spectral lines used for the calibration of the Ge and CdTe detectors.

Figure 6

Figure 5. (Color online) Quantum efficiency measurements for the Ge detector.

Figure 7

Figure 6. (Color online) The SXI quantum efficiency was measured using the average camera count per pixel for each photon of a given energy. The measurements made at X-ray energies below 8800 eV were done on the Manson. The higher-energy measurements were done on the HEX.

Figure 8

Figure 7. (Color online) Flat field image for one of the SXI cameras at the Cu 8470 eV energy.

Figure 9

Figure 8. (Color online) A flat field image using the Mg 1273 eV spectral band. It is completely uniform within 1%; the high-energy non-uniformity has completely disappeared.

Figure 10

Figure 9. (Color online) Damaged region of SXI CCD, 1273 eV.

Figure 11

Figure 10. (Color online) Quantum efficiency, front illuminated large format CCD camera.

Figure 12

Figure 11. (Color online) An image of the HEX X-ray beam from a charge injection device (CID) camera that is the sum of 50 individual images.

Figure 13

Figure 12. (Color online) Camera sensitivity for two CID cameras. The curves are models for the camera response.