Protein-functionalized WO3 nanorods–based impedimetric platform for sensitive and label-free detection of a cardiac biomarker

Published online by Cambridge University Press:  06 February 2019

Deepika Sandil
Affiliation:
Advanced Sensor Laboratory, Department of Applied Physics, Delhi Technological University, Delhi 110042, India; and Department of Applied Physics, Bhagwan Parshuram Institute of Technology, Delhi 110089, India
Suresh C. Sharma
Affiliation:
Advanced Sensor Laboratory, Department of Applied Physics, Delhi Technological University, Delhi 110042, India
Nitin K. Puri*
Affiliation:
Advanced Sensor Laboratory, Department of Applied Physics, Delhi Technological University, Delhi 110042, India
*
a)Address all correspondence to this author. e-mail: nitinkumarpuri@dtu.ac.in, nitinpuri2002@yahoo.co.in

Abstract

We report the development of a sensitive and a label-free electrochemical immunosensing platform for the detection of cardiac biomarker troponin I (cTnI) using tungsten trioxide nanorods (WO3 NRs). The low-temperature hydrothermal technique was employed for the controlled synthesis of WO3 NRs. Thin films of 3-aminopropyltriethoxy saline (APTES)-functionalized WO3 NRs were deposited on indium tin oxide (ITO)-coated glass substrate (0.5 cm × 1 cm) using electrophoretic deposition technique. The covalent immobilization of cTnI antibody onto functionalized WO3 NRs electrode was accomplished using EDC-NHS [1-(3-(dimethylamino)-propyl)-3-ethylcarbodiimide hydrochloride and N-hydroxysulfosuccinimide] chemistry. The structural and morphological characterizations of WO3 NRs and functionalized WO3 NRs were studied using X-ray diffraction, field emission scanning electron microscopy, transmission electron microscopy, Fourier transform infrared spectroscopy, and electrochemical techniques. The impedimetric response study of the proposed immunosensor demonstrates high sensitivity [6.81 KΩ mL·cm2)] in a linear detection range of 0.01–10 ng/mL. The excellent selectivity, good reproducibility, and long-term stability of the proposed immunosensing platform indicate WO3 NRs as a suitable platform for the development of a point-of-care biosensing device for cardiac detection.

Information

Type
Article
Copyright
Copyright © Materials Research Society 2019 
Figure 0

Figure 1: XRD pattern of hydrothermally synthesized nWO3.

Figure 1

Figure 2: FESEM micrograph of WO3 NRs (a) at low magnification and (b) at high magnification, and (c) selected area electron diffraction image of WO3 NRs. Inset: TEM image of WO3 NRs.

Figure 2

Figure 3: FT-IR spectra of (a) APTES/WO3 NRs/ITO electrode, (b) anti-cTnI/APTES/WO3 NRs/ITO immunoelectrode, and (c) EA/anti-cTnI/APTES/WO3 NRs/ITO immunoelectrode.

Figure 3

Figure 4: (a) Cyclic voltammogram (CV) of the different modified electrodes in PBS (pH = 7.4) at the scan rate of 50 mV/s. (b) The EIS spectra of the modified electrodes in the frequency range 0.01–105 Hz.

Figure 4

Figure 5: (a) CV response of EA/anti-cTnI/APTES/n-WO3/ITO immunoelectrode as a function of scan rate (40–160 mV/s). (b) Variation of anodic peak current (Ipa) and cathodic peak current (Ipc) with respect to square root of the scan rate. (c) Variation of redox peak potential difference with respect to square root of scan rate.

Figure 5

Figure 6: (a) EIS response study of the immunoelectrode as a function of cTnI antigen concentration (0.01–10 ng/mL); inset shows Randles equivalent circuit. (b) A calibration plot obtained between RCT and cTnI antigen concentration; inset shows sensing along with control study.

Figure 6

Figure 7: (a) Reproducibility study of the immunoelectrodes fabricated under identical conditions with cTnI (0.5 ng/mL). (b) Interference study of the immunoelectrode in the presence of different interferents. (c) Shelf life study of the EA/anti-cTnI/APTES/nWO3/ITO immunoelectrode.

Figure 7

TABLE I: Sensing characteristics of the WO3 NR–based platform compared with those reported in the literature.

Figure 8

Scheme 1: Schematic representation of fabrication of immunoelectrode for electrochemical biosensing.