Design of gripping devices based on a globoid transmission for a robotic biomaterial aliquoting system

Published online by Cambridge University Press:  10 August 2022

Artem Voloshkin
Affiliation:
Russia, Belgorod, Belgorod State Technological University named after V.G. Shukhov
Larisa Rybak*
Affiliation:
Russia, Belgorod, Belgorod State Technological University named after V.G. Shukhov
Vladislav Cherkasov
Affiliation:
Russia, Belgorod, Belgorod State Technological University named after V.G. Shukhov
Giuseppe Carbone
Affiliation:
Russia, Belgorod, Belgorod State Technological University named after V.G. Shukhov Department of Mechanical, Energy and Management Engineering, University of Calabria, Rende, 87036, Italy.
*
*Corresponding author. E-mail: rlbgtu@gmail.com
Rights & Permissions [Opens in a new window]

Abstract

Every year the shortage of biosamples is increasing, while the requirements for their quality are constantly tightening, which requires the introduction of new technological solutions. To solve these problems, a robotic system for aliquoting biological liquid was developed. The aliquotation process is described. The station includes a serial robot on which a gripper based on a globoid worm is installed. The gripping device is parameterized and takes into account the gripping force for different finger deflections. 3D models were developed using Computer Aided Design (CAD) system tools, after which working layouts were created using 3D printing. The design process and test results are discussed to show the efficiency of the built prototype with lab tests.

Information

Type
Research Article
Copyright
© The Author(s), 2022. Published by Cambridge University Press
Figure 0

Figure 1. Examples of performing aliquots.

Figure 1

Table I. Specification WR3 specs and delta robot.

Figure 2

Figure 2. Three-finger handle based on a crank-sliding mechanism.

Figure 3

Figure 3. Prototype of a gripping device.

Figure 4

Figure 4. (a) Diagram of the developed sample. 1 - Radius of rotation of the lever, 2 - lever, 3 - rotary joint, 4 - axis of rotation, 5 - spherical part of the lever, 6 - globoid worm, 7 - mechanism base, 8 - spring, 9 - position of the lever when the spring is compressed. (b) Globoid worm.

Figure 5

Figure 5. Calculation scheme of the globoid transmission.

Figure 6

Figure 6. Calculation scheme.

Figure 7

Figure 7. Creating the gripper housing. (a) Developing the flange, (b) pulling the housing, (c) developing the housing pockets.

Figure 8

Figure 8. Gripper input link, (a) flange, (b) globoid surface.

Figure 9

Figure 9. Graphic process of constructing a globoid, (a) sketch of a curved trajectory, (b) division of the globoids into parts of equal height, (c) creation of a curved surface.

Figure 10

Figure 10. 3D model of the gripper device.

Figure 11

Figure 11. Mockup motion frames, (a), (b), (c) capture motion sequence.

Figure 12

Figure 12. A robotic system for aliquoting biological material.

Figure 13

Figure 13. Moving test tubes with liquid.

Figure 14

Figure 14. Repeatability assessment.

Figure 15

Table II. Summary table of calculation parameters.

Figure 16

Figure 15. Graph of the dependence of the coefficient of elasticity on the force of the spring action.

Figure 17

Figure 16. Influence of the angles $\alpha$ and $\beta$ on the calculation system, (a) the friction force depending on the angle of rotation of the lever, (b) the friction force depending on the angle of approach of the helical surface of the globoid.

Figure 18

Figure 17. Graph of the moment at different angles $\alpha$ and $\beta$.

Figure 19

Table III. Summary table of calculation parameters.

Figure 20

Figure 18. Dependence of the required torque on the diameter of the object.