Experimental validatıon of peripheral dose distribution of electron beams for eclipse electron Monte Carlo algorithm

Published online by Cambridge University Press:  23 July 2018

Hilal Acar*
Affiliation:
Department of Radiation Oncology, Istanbul Medipol University, Istanbul, Turkey
Mustafa Caglar
Affiliation:
Department of Radiation Oncology, Istanbul Medipol University, Istanbul, Turkey
Ayse Y. Altinok
Affiliation:
Department of Radiation Oncology, Istanbul Medipol University, Istanbul, Turkey
*
Correspondence to: Assoc Prof. Hilal Acar, Department of Radiation Oncology, Medicine Faculty, Istanbul Medipol University, 34214, Istanbul, Turkey. Tel: +90 (535) 9395365. Fax: 90 (212) 460 70 70. E-mail: hacar@medipol.edu.tr
Rights & Permissions [Opens in a new window]

Abstract

Aim

The accuracy of two calculation algorithms of the Varian Eclipse treatment planning system (TPS), the electron Monte Carlo algorithm (eMC) and general Gaussian pencil beam algorithm (GGPB) for calculating peripheral dose distribution of electron beams was investigated.

Methods

Peripheral dose measurements were carried out for 6, 9, 12, 15, 18 and 22 MeV electron beams using parallel plate ionisation chamber and EBT3 film in the slab phantom. Measurements were performed for 6×6, 10×10 and 25×25 cm2 cone sizes at dmax of each energy up to 20 cm beyond the field edges. The measured and TPS calculated data were compared.

Results

The TPS underestimated the out-of-field doses. The difference between measured and calculated doses increase with the cone size. For ionisation chamber measurement, the largest deviation between calculated and measured doses is <4·29% using the eMC, but can increase up to 8·72% of the distribution using GGPB. For film measurement, the minimum gamma analysis passing rates between measured and calculated dose distributions for all field sizes and energies used in this study were 91·2 and 74·7% for eMC and GGPB, respectively.

Findings

The use of GGPB for planning large field treatments with 6 MeV could lead to inaccuracies of clinical significance.

Information

Type
Original Article
Copyright
© Cambridge University Press 2018 
Figure 0

Table 1 The dose maximum depth of each energy used in this study.

Figure 1

Table 2 Eclipse electron Monte Carlo calculation parameters used in this study.

Figure 2

Figure 1 The Roos chamber measured percentage peripheral dose distributions of 6, 12 and 22 MeV electron beams for 10×10 cm2 cone.

Figure 3

Figure 2 The Roos chamber measured percentage peripheral dose distributions of 6 MeV electron beams for 6×6, 10×10 and 25×25 cm2 cone sizes.

Figure 4

Figure 3 The Roos chamber measured percentage peripheral dose distributions of 22 MeV electron beams for 6×6, 10×10 and 25×25 cm2 cone sizes.

Figure 5

Figure 4 (a) The percentage difference between ionisation chamber measured and GGPB calculated data of 6, 9, 12, 15, 18 and 22 MeV electron beams for 6×6 cm2 cone. (b) The percentage difference between ionisation chamber measured and eMC calculated data of 6, 9, 12, 15, 18 and 22 MeV electron beams for 6×6 cm2 cone.

Figure 6

Figure 5 (a) The percentage difference between ionisation chamber measured and GGPB calculated data of 6, 9, 12, 15, 18 and 22 MeV electron beams for 10×10 cm2 cone. (b) The percentage difference between ionisation chamber measured and eMC calculated data of 6, 9, 12, 15, 18 and 22 MeV electron beams for 10×10 cm2 cone.

Figure 7

Figure 6 (a) The percentage difference between ionisation chamber measured and GGPB calculated data of 6, 9, 12, 15, 18 and 22 MeV electron beams for 25×25 cm2 cone. (b) The percentage difference between ionisation chamber measured and eMC calculated data of 6, 9, 12, 15, 18 and 22 MeV electron beams for 25×25 cm2 cone.

Figure 8

Table 3 The gamma index passing rates of in field cross plane profiles measured by ionisation chamber and film dosimetry for ±3%/3 mm criteria

Figure 9

Table 4 Percentage dose difference (film measurement−ionisation chamber measurement/film measurement) between out-of-field doses measured with EBT3 film and Roos ionisation chamber at 10×10 cone

Figure 10

Table 5 Gamma index passing rates for EBT3 film measurement and GGPB calculation for ±3%/3 gamma index passing criteria

Figure 11

Table 6 Gamma index passing rates for EBT3 film measurement and eMC calculation for ±3%/3 gamma index passing criteria

Figure 12

Table 7 Source of the uncertainty for ionisation chamber measurement

Figure 13

Table 8 Source of the uncertainty for intrinsic characteristics of EBT3 film measurement