Coarctation of the aorta-stenting via Glidesheath Slender in a newborn with recoarctation early after a Norwood operation

Published online by Cambridge University Press:  30 October 2017

Katarzyna Gendera*
Affiliation:
Department of Paediatric Cardiology and Congenital Heart Defects, German Heart Centre, Munich, Germany
Julie Cleuziou
Affiliation:
Department of Cardiovascular Surgery, German Heart Centre, Munich, Germany
Daniel Tanase
Affiliation:
Department of Paediatric Cardiology and Congenital Heart Defects, German Heart Centre, Munich, Germany
*
Correspondence to: K. Gendera, Deutsches Herzzentrum Munchen, Paediatric Cardiology and Congenital Heart Defects, Lazarettstra e 36, Munchen, 80636, Germany. E-mail: gendera.kasia@gmail.com
Rights & Permissions [Opens in a new window]

Abstract

In this article, we report on a newborn with hypoplastic left heart syndrome in whom recoarctation of the aorta was treated with a bare metal stent (Cook Formula 414 Stent) in the early postoperative period after a Norwood procedure. To reduce the risk for scarring and occluding the femoral artery the stent was implanted via 5F Glidesheath Slender sheath.

Information

Type
Brief Report
Copyright
© Cambridge University Press 2017 
Figure 0

Figure 1(a) Angiographic depiction of the aortic arch showing a narrowed segment at the level of the aortic isthmus. The peak invasive pull-back gradient was 20 mmHg. (b) After implantation of a Cook Formula 414 Stent (6×12 mm) there was no residual gradient. LPA=left pulmonary artery; re-CoA=recoarctation of the aorta.

Figure 1

Figure 2(a) The Glidesheath Slender 5F (Terumo, Tokyo, Japan). (b) In our bench test with the Cook Formula 414 Stent (6×12 mm) dilatation to a maximal diameter of 12 mm was possible. (c) The outer diameter of the Glidesheath Slender 5F is significantly smaller than the outer diameter of a regular 5F sheath. The inner lumen is compatible with 5F catheters whereas the outer diameter is similar to a regular 4F sheath. Source: Modified from Terumo Corporation, Japan.