Creation of a defined and stable Fontan fenestration with the new Occlutech Atrial Flow Regulator (AFR®)

Published online by Cambridge University Press:  04 July 2018

Anja Lehner*
Affiliation:
Department for Pediatric Cardiology and Intensive Care, Ludwig Maximilians University (LMU), Klinikum der Universität München (KUM), Munich, Germany
Ingram Schulze-Neick
Affiliation:
Department for Pediatric Cardiology and Intensive Care, Ludwig Maximilians University (LMU), Klinikum der Universität München (KUM), Munich, Germany
Nikolaus A. Haas
Affiliation:
Department for Pediatric Cardiology and Intensive Care, Ludwig Maximilians University (LMU), Klinikum der Universität München (KUM), Munich, Germany
*
Author for correspondence: A. Lehner, Department for Pediatric Cardiology and Intensive Care, Ludwig Maximilians University (LMU), Klinikum der Universität München (KUM), Marchioninistrasse 15, D-81377 Munich, Germany. Tel: +49 89 4400 73941; Fax: +49 89 4400 73943; E-mail: anja.lehner@med.uni-muenchen.de
Rights & Permissions [Opens in a new window]

Abstract

This report describes the first clinical experience with the new Occlutech Atrial Flow Regulator (AFR®) device for implementation of a late Fontan fenestration. The AFR® device secured a defined diameter of the fenestration without spontaneous re-occlusion of this extra-anatomic perforation. This ensured a permanent clinical improvement of our failing Fontan patient.

Information

Type
Brief Report
Copyright
© Cambridge University Press 2018 
Figure 0

Figure 1 Details of the AFR® device. (a) Circular shape of the device with the central hole. (b) Connection hub to the delivery cable, which is identical to the delivery set of the Occlutech Flex II® device family. (c) Flat profile of the device (here thickness 2 mm). (d) Flat profile of an implanted device (5 mm thickness).

Figure 1

Table 1 Occlutech® AFR® device measurements

Figure 2

Figure 2 Baseline anatomy. Extracardiac tunnel with a 20-mm conduit ((a) anteroposterior view, (b) lateral view). Visualisation of the right-sided pulmonary veins and the left atrium ((c) anteroposterior view, (d) lateral view). The round tip of the pig-tail catheter inside indicates the area for transseptal puncture.

Figure 3

Figure 3 Schematic illustration of the Occlutech AFR® implantation. (a) Classification of the underlying anatomy. (b) A transseptal needle is inserted from the jugular vein and the puncture is directed to the left side posteriorally to gain access to the left atrium. (c) Balloon dilatation is performed across the puncture side. (d) Final position of the Occlutech AFR® device.

Figure 4

Figure 4 Result after implantation of the Occlutech AFR® device. (a) Good alignment of the device with flat profile. (b) Right-to-left shunting (arrow).